Verify Temperature And Voltage; Saving And Restoring Camera Settings - Teledyne Linea User Manual

2k and 4k monochrome cmos line scan
Table of Contents

Advertisement

Linea 2K and 4K Monochrome Line Scan Camera User's Manual

Verify Temperature and Voltage

To d eterm ine the voltage and tem p eratu re at the cam era, u se the Refresh Voltage and Refresh
Temperature featu res fou nd in the Camera Information set.
The tem p eratu re retu rned is the internal tem p eratu re in d egrees Celsiu s. For p rop er op eration this valu e
shou ld not exceed 80 °C. If the cam era exceed s the d esignated tem p eratu re it w ill stop im aging and the
LED w ill tu rn red . Once you have d iagnosed and rem ed ied the issu e u se the reset camera fu n ction.
Note: The voltage d isp layed is the cam era's inp u t voltage. The voltage m easu rem ent featu re of the cam era
p rovid es resu lts w ithin 1% of the actu al voltage. Th e m easu rem ent can be u sed to set the ap p lied voltage
to the cam era.

Saving and Restoring Camera Settings

The p aram eters u sed to select, load and save u ser sets are grou p ed together u nd er the Cam era
Inform ation set of featu res. There are 8 u ser sets available and one factory set.
Parameter
User Set Default Selector
User Set Selector
User Set Load
User Set Save
Description of the Camera Settings
The cam era op erates in one of fou r settings:
1.
Cu rrent op eration w ith active settings.
2.
User setting.
3.
Factory setting (read -only).
4.
Defau lt setting.
The settings active d u ring the cu rrent op eration can be saved (thereby becom ing the u ser setting) u sing
the User Set Save p aram eter. A p reviou sly saved u ser setting (User Set 1 to 8) or the factory settings can
be restored u sing the User Set Selector and User Set Load p aram eters.
Either the Factory or one of the User settings can be saved as the Defau lt Setting by selecting the set in the
User Set Defau lt Selector. The chosen set au tom atically saves as the d efau lt se tting and is the set load ed
w hen the cam era is reset or p ow ered u p .
Teledyne DALSA
Camera Information
Select the cam era p aram eters to load w hen the cam era is reset or
pow ered up as the Factory set, or as User Set 1 to 8.
Selecting the set from the list autom atically saves it as the d efault
set.
Select the Factory or User set to Save or Load .
-Factory Set
-User Set 1 to 8.
Load the set specified by User Set Selector to the cam era and m ake
it the active / current set.
Save the current set as selected user set.
Choices