11.
Power Cartridge will not charge (cascade) when it is in the Eclipse,
and Eclipse is running off an AC / DC Power Source.
A.
Is the Power Cartridge Full? (YES or NO)
i.
If "YES" then the cartridge is completely charged. No further action required
ii.
If "NO" Is the external power adapter (AC or DC Adapter) connected? (YES or NO)
iii.
If "NO" then
adapter, "There are no charging capabilities at this time."
iv.
If "YES"
1.
2.
B.
Has the Eclipse recently been operating from the Power Cartridge for a prolonged period
of time or on the higher continuous liter flows (2.5 or 3.0 continuous)? (YES or NO)
C.
If "YES": The Power Cartridge may have over heated. Remove the Power Cartridge from
Eclipse and allow it to cool in a well ventilated area for approximately 20 min's, then
return the Power Cartridge to Eclipse for recharging.
D.
If "NO": There may be a problem with the Eclipses Charging System
(2) Power Cartridge or (3)External Power Charger may be faulty. "Exchange out the unit
for evaluation and possible repair".
Reasons why a Power Cartridge will not charge
•
Power Cartridge is fully charged
•
Power Cartridge is being used
•
Power Cartridge is too warm.
Troubleshooting Guide
Troubleshooting Guide
Eclipse is operating from the Power Cartridge, and not the AC or DC
Ensure that all power cords are properly connected to there receptacles.
Ensure that the LED is illuminated "Green on the AC or DC Adapter unit
(10 bars for Eclipse & 4 bars for Eclipse
(See Loss of external power,)
(remove & allow to cool for approximately 20
11436 Sorrento Valley Road • San Diego, CA 92121 •
"Your (1) Eclipse,
2)
min's)
www.sequal.com
• 858.202.3100
22
Need help?
Do you have a question about the Eclipse and is the answer not in the manual?
Questions and answers
I have a yellow light continusly on. I **** getting alarm code 8
A continuously yellow light with alarm code 8 on the SeQual Eclipse indicates a "Low O2" condition, meaning the oxygen purity is below 85% but above 70%.
This answer is automatically generated
Yellow light snd alarm sounding every few seconds
The yellow light on a SeQual Eclipse generally indicates a device issue. If an alarm sounds every few seconds along with the yellow light, it may signal a problem requiring service, such as sieve bed replacement. Additionally, specific alarm codes like AC-04 may indicate a malfunction needing customer service assistance.
This answer is automatically generated