Low Oxygen Purity Below 70%) - SeQual Eclipse Troubleshooting Manual

Hide thumbs Also See for Eclipse:
Table of Contents

Advertisement

3.

Low Oxygen Purity Below 70%)

Eclipse is:
Beeping
A.
What is the Eclipse liter flow?
B.
What Flow Mode is set?
C.
If Continuous Flow – Is Eclipse is at least 3" away from anything that may block the vents
(i.e.: drapes, bedding, fabric or upholstery, and is the cabinet filter clean?
i.
If "YES", then
instructions.
ii.
If "NO" then
D.
If Pulsing Flow – Is the respiratory rate greater than the range on the chart below:
i.
If "YES", then
then return to the pulse Mode
ii.
If "NO" then
Pulse Flow Setting
Bolus Size (+/- 15%)
1
2
3
4
5
6
Troubleshooting Guide
Troubleshooting Guide
with a
RED LED Flashing
(CONTINUOUS or PULSING)
your purity is low. Replace and repair the Eclipse. Contact your provider for further
relocate Eclipse for improved airflow and / or have your external cabinet filter cleaned.
switch to Continuous Mode until the respiratory rate is below the recommended rate,
purity is low. Exchange out the unit for evaluation and possible repair.
Min Breath Rate
16
6
32
6
48
6
64
6
80
6
96
6
11436 Sorrento Valley Road • San Diego, CA 92121 •
and the unit
IS
running.
AC Power
Supply and
Switch Setting
Power Cartridge
DC Power Supply
Max Breath
Max Breath
Rate
40
40
40
39
31
26
www.sequal.com
12
15 Amp
10 Amp
Switch Setting
DC Power Supply
Max Breath
Rate
Rate
40
40
40
40
40
40
39
31
31
25
26
20
• 858.202.3100

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the Eclipse and is the answer not in the manual?

Questions and answers

David C Morandy
February 4, 2025

I have a yellow light continusly on. I **** getting alarm code 8

Tom
February 7, 2025

Yellow light snd alarm sounding every few seconds

1 comments:
Mr. Anderson
February 10, 2025

The yellow light on a SeQual Eclipse generally indicates a device issue. If an alarm sounds every few seconds along with the yellow light, it may signal a problem requiring service, such as sieve bed replacement. Additionally, specific alarm codes like AC-04 may indicate a malfunction needing customer service assistance.

This answer is automatically generated