7.

System Malfunction

Eclipse has a
RED LED FLASHING
NOT
running
A.
To shut off the alarm, remove all power sources (external power source – AC or DC), and
the Power Cartridge, for a minimum of 15 seconds.
B.
Plug in external power (AC or DC) into the outlet, and ensure that the power cord is plugged
in the AC Adapter securely.
C.
Is the LED on top of the AC or DC Adapter lit, (YES or NO)?
i.
If "YES" go to Step D
ii.
If "NO": ensure there is power to the AC outlet.. If not, find a working outlet for Eclipse.
If the LED still is not lit, then the adapter may have a fault.
evaluation and possible repair.."
iii.
What color is the LED (RED or GREEN)?
iv.
Red – The voltage for the electrical outlet may not meet the requirements for Eclipse.
Insure there is adequate power to that outlet (e.g.; outlets that are controlled by a wall
or dimmer switch should not be used)
v.
Green – The AC or DC Adapter is working within its normal limits. Proceed on to Step
D
*NOTE: Continue to use the Eclipse, if Eclipse alarms again contact your provider for further instructions.
Troubleshooting Guide
Troubleshooting Guide
and
ALARMING
11436 Sorrento Valley Road • San Diego, CA 92121 •
CONTINUOUSLY, and the unit is
Exchange out the item for
Continued on Next Page
www.sequal.com
17
• 858.202.3100

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the Eclipse and is the answer not in the manual?

Questions and answers

David C Morandy
February 4, 2025

I have a yellow light continusly on. I **** getting alarm code 8

1 comments:
Mr. Anderson
May 14, 2025

A continuously yellow light with alarm code 8 on the SeQual Eclipse indicates a "Low O2" condition, meaning the oxygen purity is below 85% but above 70%.

This answer is automatically generated

Tom
February 7, 2025

Yellow light snd alarm sounding every few seconds

1 comments:
Mr. Anderson
February 10, 2025

The yellow light on a SeQual Eclipse generally indicates a device issue. If an alarm sounds every few seconds along with the yellow light, it may signal a problem requiring service, such as sieve bed replacement. Additionally, specific alarm codes like AC-04 may indicate a malfunction needing customer service assistance.