SECTION 12 TECHNICAL PARAMETERS
The technical parameters accessed via the "tEch" prompt at the end of user
parameters and are password controlled to prevent unauthorized access.
These parameters set the metrology for the scales. Each scale is set independently.
The parameters will set capacity, division, decimal point position, initial zero range,
auto and manual zero range as well as factory calibration.
ACTION
Enter the User Parameters section by pressing the
Pst key during the self-test when the power is turned
on. This will allow the user to set the way he wants
the scale to work by choosing specific values.
Press the U.Wt./Units key until the display shows
"tech"
From the "tech" user parameter, Press the Tare key.
The scale will ask you to enter the password.
Default Password is 0000. Enter "0" four times.
9999 will override any other user password. Press
Tare.
Select the scale to be set up by using the Local/Rem
key.
Press Tare to enter the technical section for that
scale.
Use the U.Wt./Units key to select the weighing unit
to be used for setting up the parameters for the
scale.
The arrow in the "Weight" window will indicate the
unit selected.
Press the Tare key to continue.
To display raw count, press Tare. To escape raw
count, press Tare again.
Press the U.Wt./Units key to scroll through the
menus for the chosen scale. Use Tare to select and
set parameters and the Zero key to escape without
changing data as with the user parameter section.
- 24 -
DISPLAYS
"F1 off"
"tech"
" Pin"
" Pin" " ----"
"tEch" " LocAL" " "
"tEch" "rEmotE" " "
"tEch" " Unit" " "
"Cnt"
(Cnt is raw count)
Both scales have the same
menus
except
capacity of the local scale is
fixed and cannot be entered.
"CAp". Scale capacity setting.
(in local scale you can't set
capacity)
that
the