TAKING RECTAL TEMPERATURES - RED CAPPED PROBE
1. Exchange the oral probe (blue ejection button) for the rectal probe (red
ejection button) by removing the probe from the storage channel, unplug-
ging the probe connector from the thermometer, and repeating steps 1
through 6 in the SETUP section.
2. Attach a probe cover in the same manner as in the oral procedure. Make sure
the rEC is displayed.
3. Separate the buttocks with one hand.
Using the other hand, gently insert the
probe only 1.5 cm (5/8 of an inch) inside
the rectal sphincter (less for infants and
children). The use of lubricant is optional.
WARNING: Incorrect insertion can
cause bowel perforation.
Tilt the probe so the tip is touching tissue. Keep hand separating buttocks
in place. Hold the probe in place during the entire temperature measurement
process.
4. When the final temperature is reached, a
tone will sound and the temperature will
be displayed. Rectal temperature mea-
surements will take about 15 seconds.
5. After the temperature measurement is
complete, remove the probe from the
patient, then eject the probe cover by
firmly pressing the ejection button on the probe. (Figure 1).
6. Insert the probe into the storage channel to clear the display, in preparation
for another temperature.
7. Wash hands (note: washing hands greatly reduces the risk of cross-
contamination and NSI).