Dampen a cotton swab or cotton cloth with soap solution and clean the Y axis v-groove shown below.
Slide the X-Axis Arm toward the front or rear of the device as necessary to gain access to the entire
length of the Y-axis v-groove.
Main Enclosure
Clean the glass Top/Front door with a non-abrasive cotton cloth, paper towel or facial tissue and window
cleaner. The top window is made out of glass; therefore, do not use abrasive cleaning clothes because
they will scratch the glass. Also, do not use abrasive chemicals that will damage the glass.
Use a soft cloth dampened with the soap solution to clean the enclosure. Do not use alcohol, acetone or
any other harsh chemical as these will damage the paint.

Optics

A visual inspection of the #2 and #3 mirrors, beam window and focus lens should be performed at least once
a day.
CAUTION:
Do not clean an optic that appears clean. Excessive cleaning can damage the optical
coatings. To prevent contamination, wash your hands thoroughly before handling and cleaning any optic.
Try not to touch the optical surfaces with your fingers, handle optics only by the edge or optical housing.
Fingerprints can damage the optical coatings.
Never clean any optic right after engraving or cutting because the optic may be hot and the cool lens cleaning
solution may thermally shock the optic and crack it.
#2 Mirror
To gain access to the #2 mirror rotate the X axis rail cover up and out of the way. Locate the #2 Mirror Holder
handle on the left end of the rail (red) and gently slide it out of the holder. It is held in place by magnets so you
may feel a slight resistance as you are sliding it out.