Operating Conditions

To ensure that the CoaguChek XS Meter functions
correctly, follow these guidelines:
• Use the meter at room temperature, between 59°F
and 90°F (15°C and 32°C).
• Use the meter at a relative humidity of less than
85%, without condensation.
• When testing, keep the meter level.
• If you store the meter for a period of time, remove
the batteries.
• Do not use the meter at an altitude higher than
14,000 feet (4,300 meters).
• Do not use the meter near a strong magnetic field,
such as a microwave oven, as this may interfere with
the meter's proper operation.
Note: The CoaguChek XS Meter automatically shuts off after
3 minutes if no buttons have been pushed.