Troubleshooting

Problem Solving Checklist

Problem
Monitor display is incorrect
Monitor displays "LLL" and
alarm sounds
Monitor displays event code
"Lbb" during selftesting (90207
only)
Cuff too tight
Cuff too loose
Incorrect time displayed

Servicing

There is a 12-month warranty on this product.
Field service for the ABP monitor is limited to replacing batteries and accessories. If other repair is
required, return the monitor to Spacelabs Healthcare. Contact your Spacelabs Healthcare Customer
Service Representative for shipping details.
Caution:
Any attempt to perform service or repair to the monitor will result in cancellation of the
warranty.
Possible Cause
Data not retained.
Low or no power.
May be one of the following:
time-out, no reading due to air
leak in the system, improper cuff
size, cuff not properly attached
to the monitor.
Low main battery condition.
Low backup battery condition.
Cuff placed on the patient too
tightly.
Air pump staying on too long.
Cuff placed on the patient too
loosely.
Air pump not staying on long
enough.
Batteries removed for more than
one minute (90207 only).
070-0137-03 Rev.L
Solution
Replace backup battery.
Check the batteries for a full
charge; if needed, replace or
recharge the batteries.
Isolate cause and correct.
Power OFF the monitor imme-
diately. Replace batteries within
60 seconds after removal to
continue monitoring.
Replace backup battery before
continuing.
Reposition the cuff.
Return monitor to Spacelabs
Healthcare for service.
Reposition the cuff.
Return monitor to Spacelabs
Healthcare for service.
Re-initialize monitor.