Setup Test
Note:
Verify that the cable connections are secure.
Power ON the ABP monitor. It will display "9999." While the monitor is being read or initialized, the digits
will change to indicate that communication is taking place between the monitor and the analysis system.
When communication is complete, the digits will no longer change.
Offi ce Check Mode
The monitor automatically enters an Offi ce Check mode for the fi rst fi ve measurements immediately
following initialization. This allows you to verify the performance of the monitor on an individual patient
without the need for re-initialization to reset the display features.
While in the Offi ce Check mode, the monitor operates as follows:
Displays the cuff pressure on each bleed step.
Displays systole, diastole, and heart rate at the end of the measurement.
Bleeds one step below the diastolic value as determined by the monitor.
Terminating Offi ce Check Mode
For the 90207 monitor (versions earlier than 2.14), press the START/STOP key twice to cancel each of
the remaining readings. The Offi ce Check mode is terminated when the sum of the cancelled and suc-
cessful measurements equals fi ve.
For the 90217 and the 90207 monitors (versions 2.14 and later), press the START/STOP key twice to
cancel a single blood pressure reading. The Offi ce Check mode is terminated when a blood pressure
reading is cancelled.
Any event that prevents a successful blood pressure measurement (other than a manual cancel) is not
counted as one of the fi ve Offi ce Check mode readings.
Reinstating Offi ce Check Mode
The Offi ce Check mode may be reinstated in the 90217 and the 90207 monitors (versions 2.14 and
later).
To reinstate the Offi ce Check mode without initialization of the monitor:
1.
Power ON the monitor.
2.
Press and hold the START/STOP key when the version displays on the LCD.
3.
Release the START/STOP key when EC03 displays on the LCD.
An EC13 will be logged to indicate the time at which the Offi ce Check was reinstated. The Offi ce Check
mode will be enabled for fi ve additional successful measurements.
1-6
Operation
90207/90217 Operations Manual