Operation
Diastolic pressure value cannot be obtained from the data available.
90207
Systolic pressure cannot be obtained from the data available.
90217
Algorithm could not process input data quickly enough, resulting in an
input queue overfl ow.
90207
Mean arterial pressure cannot be obtained from the data available.
90217
This monitor must be initialized.
Heart rate value cannot be obtained from the data available.
Heart rate value cannot be obtained from the data available.
Bleed steps were too small. This may be caused by a partially obstructed
air hose. All blood pressure attempts following this message are inhibited.
Attempts can be enabled by powering the monitor OFF and then ON.
Unexpected or contradictory data (such as a negative cuff pressure).
Low battery detected prior to start of measurement.
Low battery detected after measurement started. Usually caused by the
pump drawing enough current to lower the battery voltage.
The report does not print an event code for this condition, which is a low
backup battery. Contact Spacelabs Medical for replacement of the bat-
tery.
90207/90217 Operations Manual