Commissioning
7.3 Description of OSD menu
Service level 2
Others
54
Function
Reset to factory defaults
Restart
Info
Serial interface
Serial bus
Adjustment/range
Execute
Execute
On
Off
Default: Off
On
Off
Default: On
Enable
Disabled
Default: Disabled
Instruction Manual, 05/2008, A5E01588521A-002
Description
All parameters are reset
The following are deleted:
User settings
Custom settings
The following are reset:
Backlight
Sharpness
User color location
Signal settings
Black level settings
RGB relation
DPMS settings
Language
OSD setting
Gamma curve
Standard group
Info window settings
Serial interface settings
H-Scaler clip
The processor is reset
The power supply unit is not
switched off. The monitor
restarts.
To display an info window
The info window is displayed
on the bottom right of the
screen if the source
changes. It provides more
detailed information about
the set timing.
The serial interface can be
activated or deactivated.
If the serial interface is
deactivated ("Off"), it is
possible to suppress e.g. the
downloading of firmware.
To configure a network of
several displays
This function is used to
interconnect several displays
so that they can all be
calibrated using a single
computer (connected to the
first display).
Note:
See the section "Serial
interface" (Page 32)