AirSep Focus Portable Patient Manual page 30

Portable oxygen concentrator
Hide thumbs Also See for Focus Portable:
Table of Contents

Advertisement

Problem
Unit alarms, does not start in automobile while
connected to a properly functioning DC outlet.
Alarm condition,
Audible alarm sounds
and the (Red) alarm light illuminates.
Solid (Amber) light without audible alarm.
All other problems.
MN170-1 rev A 03/11
Probable Cause
Focus power supply needs resetting.
A general malfunction has occurred.
Focus need inspection and / or service.
Solution
Turn off unit. Disconnect the DC power supply from
the automobile outlet, restart the automobile, and
then reconnect the DC power supply into the
automobile DC outlet to reset the DC power supply.
Turn off unit. Change to another source of oxygen as
available, and contact your Equipment Provider.
Contact your Equipment Provider. It is time for your
Focus unit to be inspected and / or serviced by your
Equipment Provider. After any necessary service and
the performance is verified by the Equipment
Provider, the alarm service/indicator light will be
reset.
Turn off unit. Change to another source of oxygen as
available, and contact your Equipment Provider.