[Read the Important Safety Rules section before operating this equipment]

Audible Alarm and Indicator Lights

When the Focus unit senses inhalation, a pulse of oxygen is delivered through the nasal cannula. The green light on the unit's
control panel blinks each time a breath is detected.
Additionally, when the unit is ope rating and the battery is simultaneously being charged through the AC or DC power s upply,
the battery for the Focus unit will displa
gauge/indicator(s) and remain on for approximately one hour after reaching a full charge.
An audible alarm sounds if Focus has a low battery, if the cannul a is disconnected, or if perfor mance of the unit is outside of
specifications. The light and audi ble alarm conditions are explaine d in detail below and summ arized on the chart later in this
section of the manual.
 Start-Up
A brief alarm sounds and the green and red LEDs (lights) will alternate at start-up. Focus begins to operate when the
alternating LEDs (light) stop and green LED (lights) remains on.
 Low Battery
 Focus unit indicators: As the battery power approaches a low level, the amber light on the Focus unit will flash on for
½ second with a 5-second pa use and simultaneously, a ½ second alarm sounds with a 5-second pause. Following the
battery warning indicators, if action is not taken, the unit w ill shut down. This will b e indicated when the a mber light
flashes 2 times with a 5-second pause and simultaneously, a ½ second alarm sounds 2 times with a 5 second pause.
 Battery indicator: The green light indicator on the battery gauge (Figure 9) light illuminates intermittently.
When either of these conditions occurs, connect Focus to a DC power outlet or to an AC power outlet, or change to
another source of oxygen within tw o minutes. The level of batte ry charge is indicated by th e battery gauge/indicator(s).
You can also check the state of charge at any time by pressing the button.
As noted above, when the unit is connected to AC or DC power outlet, you may simultaneously charge the Focus battery
supplied with the unit while using the unit.
In the event of an alarm or you observe Focus is not working properly; consult the Troubleshooting section in this
manual. If you cannot resolve the problem, consult your Equipment Provider.
MN170-1 rev A 03/11
y the charge level of the battery (25%
to 100% state of ch arge) on the battery
20