CALIBRATION

All Hanna Instruments thermometers have been
accurately pre-calibrated at the factory.
It is generally recommended to have all ther-
mometers recalibrated at least once a year.
For an accurate annual recalibration, contact
your nearest Hanna Service Center.
Hanna Test Plugs provide a quick and easy
way to test the meter's accuracy by simply
connecting these Test Plugs to the probe
connector of the meter. If the reading differs
by more than 0.4°C (0.8°F) from the Test
Plug value, the unit is due for recalibration.
Choose the right Test Plug to suit your needs:
For HI 9060 and HI 9061:
HI 765-18C Calibration key,-18.0°C ±0.4°C
HI 765000C Calibration key, 0.0°C ±0.4°C
HI 765070C Calibration key,+70.0°C±0.4°C
HI 765-004F Calibration key, -0.4°F ±0.8°F
HI 765032F Calibration key,+32.0°F ±0.8°F
HI 765158F Calibration key,+158.0°F±0.8°F
For HI 91541:
HI 764-18C Calibration key,-18.0°C ±0.4°C
HI 764000C Calibration key, 0.0°C ±0.4°C
HI 764070C Calibration key,+70.0°C±0.4°C
HI 764-004F Calibration key, -0.4°F ±0.8°F
HI 764032F Calibration key,+32.0°F ±0.8°F
HI 764158F Calibration key,+158.0°F±0.8°F