Getting Ready; Cleaning Items - Branson DHA-1000 Operator's Manual

Tabletop ultrasonic cleaner
Table of Contents

Advertisement

Getting Ready

Step
Action
1
Select your cleaning chemistry (check with your chemical
supplier for solution effects on metals).
2
Allowing for the volume of the parts you will be cleaning and
cleaning chemistry, fill the tank with warm tap water to the
operating level (two to four inches from the top).
3
Add cleaning chemistry to the tank water.
4
Plug the cleaner into a grounded outlet.
5
For maximum efficiency, refer to page 9, "Optimizing Your
Cleaner" before proceeding.
If this is the first time you are running the cleaner, or if you have changed
cleaning solution, you must degas the solution. This is done by setting
the cleaner up for operation and allowing the ultrasonics to drive the
warm solution for 5-10 minutes This will force out excess gas and assure
optimum cleaning. You will likely notice a change in the sound of the unit
as is degasses.

Cleaning Items

Step
Action
1
Place the items to be cleaned into a basket or other carrier.
2
Slowly lower the basket into the tank. Do not allow items to
contact the tank bottom. Do not stir the solution.
3
When items are clean, slowly remove them from the cleaner.
4
Rinse the clean items with fresh water and dry them, if
necessary.
NOTE:
8