NE-C29-C1_D_M.qxd 11.11.24 4:16 PM Page 2 Before Using the Unit INTENDED USE Medical Purpose This product is intended to be used for inhaling medication for respiratory disorders. Intended User • Legally certified medical experts, such as doctor, nurse and therapist, or healthcare personnel or patient under the guidance of qualified medical experts.
NE-C29-C1_D_M.qxd 11.11.24 4:16 PM Page 3 IMPORTANT SAFETY NOTES To assure the correct use of the product basic safety measures should always be followed including the precautions listed below: WARNING: Indicates a potentially hazardous situation which, if not avoided, could result in death or serious injury.
Page 5
Do not use or store the device while the air tube is bent. Approved for human use only. Use only original OMRON parts and accessories. Use of unsupported parts may damage the device. Do not add more than 7 ml of medication to the medication cup.
Page 6
NE-C29-C1_D_M.qxd 11.11.24 4:16 PM Page 5 IMPORTANT SAFETY NOTES Never place, or attempt to dry the compressor, the power cord or nebulizing parts in a microwave oven. General Safety Precautions: Inspect the device and parts before using the them each time, and check that there are no problems.
NE-C29-C1_D_M.qxd 11.11.24 4:16 PM Page 8 GUIDE TO USING THE NE-C29 HOW TO PREPARE THE NEBULIZER FOR USE (Page 9) NOTE: Clean and disinfect the nebulizer kit and applicable inhalation attachments before using the device for the first time, or if it has not been used for a while.
NE-C29-C1_D_M.qxd 11.11.24 4:16 PM Page 9 Operating Instructions HOW TO PREPARE THE NEBULIZER FOR USE WARNING: Clean and disinfect the nebulizer kit, mouthpiece and optional nosepiece or masks before using them for the first time after purchase, if the device has not been used for a long period of time.
Page 11
NE-C29-C1_D_M.qxd 11.11.24 4:16 PM Page 10 HOW TO PREPARE THE NEBULIZER FOR USE 5. Remove the vaporiser head from the medication cup. 6. Add the correct amount of prescribed medication to the medication cup. CAUTION: The capacity of the medication cup is 2 to 7 ml.
Page 12
NE-C29-C1_D_M.qxd 11.11.24 4:16 PM Page 11 HOW TO PREPARE THE NEBULIZER FOR USE 9. Attach the desired inhalation attachment, as described below. Attach the cap when using the mouthpiece or optional nosepiece. Using the cap with these attachments will reduce the amount of medication that is discharged into the air.
Page 13
NE-C29-C1_D_M.qxd 11.11.24 4:16 PM Page 12 HOW TO PREPARE THE NEBULIZER FOR USE 10. Attach the air tube. 1) Attach the air tube to the air connector on the compressor. 2) Attach the other end of the air tube to the air tube connector on the medication cup.
NE-C29-C1_D_M.qxd 11.11.24 4:16 PM Page 13 HOW TO TAKE MEDICATION 1. Hold the nebulizer kit as indicated on the right. CAUTION: Do not tilt the nebulizer kit so that it is at an angle of greater than 45°. Medication may flow into the mouth.
Page 15
NE-C29-C1_D_M.qxd 11.11.24 4:16 PM Page 14 HOW TO TAKE MEDICATION 3. Inhale medication as according to your doctor’s and/or health care adviser’s instructions. Using the Mouthpiece Insert the mouthpiece into your mouth and inhale the medication breathing normally. Exhale as normal through the mouthpiece.
Page 16
NE-C29-C1_D_M.qxd 11.11.24 4:16 PM Page 15 HOW TO TAKE MEDICATION 4. Complete your treatment. 1) When you have finished inhaling your medication, turn the compressor off. Press the power switch to the off position. 2) Unplug the air tube from the nebulizer kit.
NE-C29-C1_D_M.qxd 11.11.24 4:16 PM Page 16 Care and Maintenance CLEANING AND DAILY DISINFECTING Following the cleaning instructions after each use will prevent any remaining medication in the bottle from drying resulting in the device not nebulizing effectively and also for preventing infections.
NE-C29-C1_D_M.qxd 11.11.24 4:16 PM Page 17 MAINTENANCE AND STORAGE 1. Assemble the nebulizing parts. 2. If the compressor is dirty, wipe it clean with a soft cloth moistened with water or mild detergent. WARNING: The compressor is not waterproof. 3. If the outside of the air tube is dirty, wipe it clean with a soft cloth moistened with water or mild detergent.
Page 19
NE-C29-C1_D_M.qxd 11.11.24 4:16 PM Page 18 MAINTENANCE AND STORAGE To keep your device in the best condition and protect the device from damage follow these directions: CAUTION: • Do not drop the compressor or nebulizer kit, or subject them to strong shocks.
Do not boil the air filter cover. Remove the air filter cover, wash it, and make sure it is dry and reattached correctly. • Please contact your OMRON dealer for additional air filters, or if the parts supplied with the device are damaged or missing.
• If the suggested remedy does not solve the problem, do not try to repair the device - no parts of the unit are user serviceable. Return the unit to an authorized OMRON dealer. • The device should be serviced once a year by an approved OMRON service centre.
• This unit fulfils the provisions of the EC directive 93/42/EEC (Medical Device Directive) and the European Standard EN13544-1:2007, Respiratory therapy equipment - Part1: Nebulizing systems and their components. • This OMRON product is manufactured under the strict quality system of OMRON HEALTHCARE Co., Ltd., Japan. = Class II...
Page 23
NE-C29-C1_D_M.qxd 11.11.24 4:16 PM Page 22 SPECIFICATIONS Technical data for the OMRON Comp Air Pro compressor with the OMRON V.V.T. (Virtual Valve Technology) Nebulizer kit: Particle Size: *MMAD **approximately 3µm Medication Tank Capacity: 7 ml maximum Appropriate Medication Quantities: 2 ml minimum - 7 ml maximum...
Page 24
This standard defines the levels of immunity to electromagnetic interferences as well as maximum levels of electromagnetic emissions for medical devices. This medical device manufactured by OMRON HEALTHCARE conforms to this EN60601-1-2:2007 standard for both immunity and emissions. Nevertheless, special precautions need to be observed: •...
Page 25
NE-C29-C1_D_M.qxd 11.11.24 4:16 PM Page 24 SPECIFICATIONS Correct Disposal of This Product (Waste Electrical & Electronic Equipment) This marking shown on the product or its literature, indicates that it should not be disposed of, with other household wastes at the end of its working life. To prevent...
NE-C29-C1_D_M.qxd 11.11.24 4:16 PM Page 25 OPTIONAL PARTS AND ACCESSORIES Nebulizer Kit Set Nosepiece Model No. C28NEBKIT Model No. C28NP Nebulizer Kit Mouthpiece Air Tube Mouthpiece Filter Cover Model No. C30TUBE Model No. C28MP Model No. C28FLCVR Air Filters (x 5)
Page 27
Compressor Nebulizer System Model NE-C29...
Page 28
!" !"# KKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKK O ! KKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKK Q ! KKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKK S kbJ`OV= !"#$%&' KKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKK T !"# KKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKK NN !" !"#$ KKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKK NQ !" KKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKK NR !"#$%& KKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKK NS !"#$% KKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKK NT KKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKK NU !"#! KKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKKK NV...