  • Page 1 INSTRUCTION MANUAL Compressor Nebulizer NE-C801 Caution: Federal (USA) law restricts this device to sale by or on the order of a physician or licensed healthcare practitioner...
  • Page 2: Table Of Contents

    CONTENTS Before Using the Device Introduction ............3 Intended Use .
  • Page 3: Before Using The Device

    Before Using the Device INTROdUCTION Thank you for purchasing the OMRON NE-C801 Compressor Nebulizer. Fill in for future reference. DATE PURCHASED: __________________ SERIAL NUMBER: ____________________ • Staple your purchase receipt here • Register your product on-line at www.register-omron.com This product was developed for the successful treatment of asthma, allergies and other respiratory disorders.
  • Page 4: Intended Use

    Intended Use The NE-C801 Compressor Nebulizer System is intended to provide air to the pneumatic nebulizer in order to aerosolize medications for inhalation by the patient for respiratory disorders. The system is designed for use with pediatric (defined by the prescribed medication) and adult patients in the home, hospital, and sub-acute settings.
  • Page 5: Safety Information

    SAfETy INfORMATION To ensure the correct use of the product, basic safety measures should always be followed including the warnings and cautions listed in this instruction manual. SaFeTY SYMBOLS USeD In ThIS InSTrUCTIOn ManUaL Indicates a potentially hazardous situation which, if not WARNING avoided, could result in death or serious injury.
  • Page 6 SAfETy INfORMATION OperaTIng The DevICe (continued) Do not operate where oxygen is being administered in a closed environment (such as an oxygen tent). Provide close supervision when this device is used by, on, or near infants, children or compromised individuals. If the device is used continuously, the service life of the device may be shortened.
  • Page 7: Risk Of Electrical Shock

    Do not use or store the device in humid locations, such as a bathroom. Use the device within the operating temperature and humidity. Use only the AC adapter designed by Omron for this device. Use of any other AC adapter may damage the device.
  • Page 8: Maintenance And Storage

    SAfETy INfORMATION rISk OF eLeCTrICaL ShOCk (continued) Do not overload power outlets. Plug the AC adapter into the appropriate voltage outlet. Do not use extension cords. Plug the AC adapter directly into the electrical outlet. Unplug the AC adapter from the electrical outlet after using the device. Remove the AC adapter from the device after use.
  • Page 9: Know Your Unit

    Components and Replacement Parts KNOW yOUR UNIT Nebulizer Kit Compressor (Main Unit) Nebulizer Kit Holder Air Tube Plug Air Tube Connector Air Tube Plug Power Switch Back View Air Tube (PVC, 100 cm) Only use the NE-C801 Nebulizer Kit with this NOTE: Air Filter Cover Power...
  • Page 10 KNOW yOUR UNIT Components: Air Filters Instruction Manual Storage Bag Package of 5 INSTRUCTION MANUAL Compressor Nebulizer NE-C801 Caution: Federal (USA) law restricts this device to sale by or on the order of a physician or licensed healthcare practitioner AC Adapter Terminal Power Cord Power Plug...
  • Page 11: Operating Instructions

    Operating Instructions PREPARING THE NEbULIzER fOR USE WarnIng • Clean and disinfect the nebulizer kit, mouthpiece, and optional nosepiece or optional masks before using them for the first time after purchase. • If the device has not been used for a long period of time, or if more than one person uses the same device, clean and disinfect the nebulizer kit, mouthpiece, and optional nosepiece or optional masks before using them.
  • Page 12 PREPARING THE NEbULIzER fOR USE R otate the inhalation top counterclockwise (A) and lift (B) to remove the inhalation top from the medication cup. Add the correct amount of prescribed medication to the medication cup. WarnIng • For type, dose, and regime of medication follow the instructions of your physician or licensed healthcare practitioner. •...
  • Page 13 PREPARING THE NEbULIzER fOR USE Attach the desired inhalation accessory. USIng The MOUThpIeCe Place the cap on the medication cup over the inhalation air inlet (A). Attach the mouthpiece (B). nOTe: Place the cap on the medication cup when using the mouthpiece. Using the cap with the mouthpiece will reduce the amount of medication that is discharged into the air.
  • Page 14: Attaching The Air Tube

    ATTACHING THE AIR TUbE Twist the air tube plug slightly and push it firmly into the air tube connector on the upper side of the compressor (A). Twist the air tube plug slightly and push it firmly into the air tube connector on the bottom of the nebulizer kit (B).
  • Page 15: Using The Device

    USING THE dEVICE WarnIng • Do not use the device where the device may be exposed to flammable gas or vapors. • Do not use or store the device in humid locations, such as a bathroom. Use the device within the operating temperature and humidity. CaUTIOn •...
  • Page 16 USING THE dEVICE CaUTIOn • Do not block the air filter cover. • Limit use to 20 minutes at a time, and allow a 40 minutes interval before using the device again. • When using the device, the main unit may become hot. •...
  • Page 17 USING THE dEVICE Turn the power switch to the off ( ) position. The compressor turns off and nebulization stops. Disconnect the air tube from the nebulizer kit. Hold the air tube plug and gently pull down to disconnect the air tube plug from the air tube connector on the bottom of the nebulizer kit.
  • Page 18: Care And Maintenance

    Care and Maintenance CLEANING AfTER EACH USE Following cleaning instructions after each use will prevent any remaining medication in the bottle from drying resulting in the device not nebulizing effectively and will help prevent infections. WarnIng Wash the nebulizer parts after each use. Dry the parts immediately after washing. Remove the inhalation accessory (mouthpiece, optional masks, or optional nosepiece) from the nebulizer kit.
  • Page 19: Daily Disinfecting

    dAILy dISINfECTING Disinfect the nebulizer kit and mouthpiece, or optional masks or optional nosepiece after the last treatment of the day. To disinfect the parts, use one of the methods described below: A. Use a commercially available disinfectant. Follow the instructions provided by the disinfectant manufacturer.
  • Page 20: Changing The Air Filter

    CaUTIOn Before inserting the new air filter make sure the air filter is clean and free of dust. Do not operate the device without the air filter. Use only the Omron air filter designed for this device. Put the air filter cover back...
  • Page 21: Caring For The Device

    • Do not subject the compressor or any of the components to strong shocks, such as dropping on the floor. • Use only Omron authorized parts and accessories. Parts and accessories not approved for use with the device do not perform the expected specification or it may damage the unit.
  • Page 22: Troubleshooting

    Troubleshooting TROUbLESHOOTING GUIdE CaUSe SOLUTIOn prOBLeM Turn the power switch off. The AC adapter is not No power to unit when Plug the AC adapter into an plugged into an electrical the power switch is on. electrical outlet. Turn the device outlet.
  • Page 23: Limited Warranty

    TERMS THAT MAY BE IMPOSED BY LAW, IF ANY, ARE LIMITED IN DURATION TO THE PERIOD OF THE ABOVE EXPRESS WARRANTY. OMRON SHALL NOT BE LIABLE FOR LOSS OF USE OR ANY OTHER SPECIAL, INCIDENTAL, CONSEQUENTIAL OR INDIRECT COSTS, EXPENSES OR DAMAGES.
  • Page 24: Fcc Statement

    fCC STATEMENT nOTe: POTENTIAL FOR RADIO/TELEVISION INTERFERENCE (for U.S.A. only) This product has been tested and found to comply with the limits for a Class B digital device, pursuant to part 15 of the FCC rules. These limits are designed to provide reasonable protection against harmful interference in a residential installation. The product generates, uses, and can radiate radio frequency energy and, if not installed and used in accordance with the instructions, may cause harmful interference to radio communications.
  • Page 25: Specifications

    SPECIfICATIONS NE-C801 (NE-C801S-Z) Model: Compressor nebulizer Type: 100 - 240V ~ 350mA, 50/60 Hz Rating (AC adapter): 12V 0.8A Rating (Compressor Nebulizer): Approx. 0.3 mL/min (without cap) Nebulization Rate: *MMAD approx. 3 μm (based on EN13544-1:2007) Particle Size: 7 mL (cc) max. Medication Cup Capacity: 2 mL - 7 mL (cc) Appropriate Medication Quantities: Intermittent operation 20 min. ON/40 min. OFF...
  • Page 26: Technical Data

    Technical DaTa Particle Specifications A series of aerosol performance tests were performed using an 8 stage cascade impactor at a sampling flow rate of 15 l/min equipped with a USP <601> induction port throat. Aerosol was sampled directly from the outlet. Three (3) device samples were tested with 3 runs each, for a total of 9 sample points per each drug.
  • Page 27: Guidance And Manufacturer's Declaration

    Nevertheless, special precautions need to be observed: • The use of accessories and cables other than those specified by OMRON, with the exception of cables sold by OMRON as replacement parts for internal components, may result in increased emission or decreased immunity of the device. • The medical devices should not be used adjacent to or stacked with other equipment.
  • Page 28 GUIdANCE ANd MANUfACTURER'S dECLARATION Electromagnetic immunity IEC60601-1-2 IEC 60601 Compliance Electromagnetic environment – Immunity test Test level level guidance Floor should be wood, concrete, or Electrostatic ±6 kV ±6 kV ceramic tile. If floors are covered discharge (ESD) contact contact with synthetic material, the relative IEC 61000-4-2 ±8 kV air ±8 kV air...
  • Page 29 GUIdANCE ANd MANUfACTURER'S dECLARATION Electromagnetic immunity IEC60601-1-2 IEC 60601 Compliance Electromagnetic environment – Immunity test Test level level guidance Portable and mobile RF communications equipment should be used no closer to any part of the NE-C801 including cables, than the recommended separation distance calculated from the equation appropriate to the frequency of the transmitter.
  • Page 30 GUIdANCE ANd MANUfACTURER'S dECLARATION Recommended separation distance between portable and mobile RF communications equipment and the NE-C801 The NE-C801 is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The customers or the users of the NE-C801 can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF communications equipment (transmitters) and the NE-C801 as recommended below, according to the maximum output power of the communications equipment.
  • Page 32 Distributed by: OmrOn HealtHcare, Inc. 1925 West Field Court Lake Forest, IL 60045 U.S.A. www.omronhealthcare.com toll-free: 1-800-634-4350 Made in China © 2011 OMRON HEALTHCARE, INC.