Setting Up Ifc Custom Programme - Verity Medical NeuroTrac IFC Rehab Operation Manual

Ifc stimulator/dual channel tens&nms stimulator
Table of Contents

Advertisement

5.2. Setting up IFC Custom programme.

Continuous mode custom IFC programme (PC1)
1. Select PC1 by pressing IFC PROG. button
2. Press and hold IFC SET/ESC button for 3 seconds to enter custom pro�
gramme parameter adjustment menu
3. Flashing Hz symbol indicates that frequency can be adjusted. Use "CH2 +"
or "CH2 -" to adjust frequency in range from 1 � 150 Hz ( 1 Hz step)
4. Once frequency is adjusted press SET/ESC button to save settings and
return to Home screen
Work/Rest mode IFC custom programme(PC2)
1. Select PC2 by pressing IFC PROG. button
2. Press and hold SET/ESC button for 3 seconds to enter custom programme
parameter adjustment menu
3. Flashing Hz symbol indicates that frequency can be adjusted. Use "CH2 +"
or "CH2 - " button to adjust frequency in range from 1 � 150 Hz ( 1 Hz step)
4. Press "CH1+" button to change to work time adjustment screen. Flashing
WORK segment indicates that work period can be adjusted. Use "CH2 +"
or "CH2 - " button to adjust the time in range from 1 � 99 seconds (1 sec.
increments).
5. Press "CH1+" button again to change to rest time adjustment screen.
Flashing REST segment indicates that rest period can be adjusted. Use
"CH2 +" or "CH2 - " button to adjust the time in range from 1 � 99 seconds
(1 sec. increments).
6. Press SET/ESC button to save settings and return to home screen.
If "CH1+" button was pressed again rather than SET/ESC then the device
would loop back to the Hz adjustment screen, similarly if "CH1�" button
was pushed LCD would loop back to Work time setting screen.
7. Repeat above procedure to re�adjust custom programme.
NeuroTrac™ IFC Rehab Operation Manual
12

Advertisement

Table of Contents
loading