Replacing the Trooper Scale Load Cell Assembly ............. 3-4 Replacing the Trooper Count Scale Membrane Switch ..........3-6 Replacing the Trooper Count Scale Main PC Board ..........3-7 Replacing the Trooper Count Scale Display PC Board ..........3-8 Replacing the Trooper Count Scale Load Cell Assembly..........3-9...
Page 6
Installation ...................... B-1 B.2.1 Minimum Requirements ................B-1 B.2.2 Connecting the RS232 Interface ..............B-1 B.2.3 Installing the Trooper Count Scale Tools Software Program ......B-1 B.2.4 Opening the Trooper Count Scale Tools Software Program ......B-1 Screen Operations ....................B-2 B.3.1 Configuration (Config) ................
Page 7
Trooper Scale - Cannot Calibrate............2-4 Trooper Scale - Communications Failure ..........2-6 Trooper Scale Error Codes ..............2-8 Trooper Count Scale - Will Not Turn on With AC Adapter ......2-9 Trooper Count Scale - Will Not Turn on Using Batteries ......2-10 2-10 Trooper Count Scale - Does Not Respond to Front Panel Controls ....
Page 8
LIST OF ILLUSTRATIONS (Cont.) FIGURE NO. TITLE PAGE Sample Calculation of Standard Deviation ..........4-4 Off-Center Load Test Weight Locations ..........4-4 RS232 Interface Pin Connections ............4-8 Expoded View of Trooper Scale and Trooper Count Scale......5-2...
Chapter 3 Repair Procedures - Contains detailed repair procedures for all major components. Chapter 4 Testing - Contains performance tests, specifications and communications testing procedures. Chapter 5 Parts Lists - Contains exploded views of Trooper Scale and Trooper Count Scale identifying all serviceable replacement components with associated parts lists.
CALIBRATION MASSES REQUIRED The masses required to test the Ohaus Trooper Scale and Trooper Count Scale must meet the requirements of ASTM Class 4 or OIML F2. The mass values are listed in Table 1-1. Bolded values are default settings.
CHAPTER 1 INTRODUCTION 1.4 SERVICE STRATEGY The repair method for the Trooper Scale and Trooper Count Scale is the direct substitution of major assemblies. The Trooper Scale contains 3 major replaceable assemblies: Cover Assembly with Membrane Switch, Load Cell Assembly with Frame and the Main PC Board with LCD.
Trooper Count Scales are covered in Tables 2-8 through 2-14. The troubleshooting tables refer to interconnection and wiring diagrams in this section to assist in locating the problem. TABLE 2-1 TROOPER SCALE - WILL NOT TURN ON WITH AC ADAPTER. REMEDY SYMPTOM PROBABLE CAUSE Check the ac adapter voltage output.
CHAPTER 2 DIAGNOSIS TABLE 2-2 TROOPER SCALE - WILL NOT TURN ON USING BATTERIES. PROBABLE CAUSE REMEDY SYMPTOM Check position of batteries. Incorrect battery installation. Scale will not turn on with new batteries installed. Remove the 6 screws from the bottom...
CHAPTER 2 DIAGNOSIS TABLE 2-3 TROOPER SCALE - DOES NOT RESPOND TO FRONT PANEL CONTROLS. PROBABLE CAUSE SYMPTOM REMEDY 1. Open the scale, remove six screws Membrane Switch is defec- Scale does not respond to front from the bottom of the scale.
CHAPTER 2 DIAGNOSIS TABLE 2-4 TROOPER SCALE - NO DISPLAY OR PARTIAL DISPLAY. PROBABLE CAUSE REMEDY SYMPTOM The Main PC Board is replaced as a Main PC Board is defective. Display is not on or partial whole assembly. Check procedures in characters are displayed.
Page 17
CHAPTER 2 DIAGNOSIS TABLE 2-5 TROOPER SCALE - CANNOT CALIBRATE (Cont.). SYMPTOM PROBABLE CAUSE REMEDY Scale can be turned on but will Load Cell assembly defec- With the platform in place and the scale not calibrate (Cont.). tive. turned on, measure across pins -SIG and +SIG on the PC Board.
CHAPTER 2 DIAGNOSIS TABLE 2-6 TROOPER SCALE - COMMUNICATIONS FAILURE. SYMPTOM PROBABLE CAUSE REMEDY Communications cannot be Communication cable is not Tighten connectors and use the proper established with scale turned plugged in or the incorrect cable. cable is being used.
Main PC Board Black TARE ON/ZERO & OFF Battery Holder G/N/T & MENU PRINT UNITS COMM Black NOT USED Figure 2-1. Trooper Scale Interconnection Diagram. Print G/N/T On/Zero Tare Units Menu MEMBRANE SWITCH Figure 2-2. Trooper Scale Membrane Switch Wiring Diagram.
CHAPTER 2 DIAGNOSIS 2.5 TROOPER SCALE ERROR CODE TABLE The Trooper Scale is equipped with software which will display an error condition when it occurs. Review the listed codes and follow instructions to correct the problem. TABLE 2-7. TROOPER SCALE ERROR CODES...
CHAPTER 2 DIAGNOSIS TABLE 2-8 TROOPER COUNT SCALE - WILL NOT TURN ON WITH AC ADAPTER. SYMPTOM PROBABLE CAUSE REMEDY Scale will not turn on with AC Adapter defective. Check the ac adapter voltage output. adapter supplied. The ac adapter output should measure 12-16 Volts dc.
CHAPTER 2 DIAGNOSIS TABLE 2-9 TROOPER COUNT SCALE - WILL NOT TURN ON USING BATTERIES. PROBABLE CAUSE REMEDY SYMPTOM Incorrect battery installation. Check position of batteries. Scale will not turn on with new batteries installed. Wiring harness defective or Remove the 6 screws from the bottom...
CHAPTER 2 DIAGNOSIS TABLE 2-10 TROOPER COUNT SCALE - DOES NOT RESPOND TO FRONT PANEL CONTROLS. PROBABLE CAUSE SYMPTOM REMEDY 1. Open the scale, remove six screws Membrane switch defective. Scale does not respond to front from the bottom of the scale.
CHAPTER 2 DIAGNOSIS TABLE 2-11 TROOPER COUNT SCALE - NO DISPLAY OR PARTIAL DISPLAY. SYMPTOM PROBABLE CAUSE REMEDY Display is not on or partial Either the display board con- Remove the 6 screws from the bottom characters are displayed. nector is loose from the Main...
CHAPTER 2 DIAGNOSIS TABLE 2-12 TROOPER COUNT SCALE - CANNOT CALIBRATE. PROBABLE CAUSE SYMPTOM REMEDY Incorrect weights. Verify that proper weights are used. Scale can be turned on but will not calibrate. The software lock menu is Set the software lock menu to OFF. Ac- set ON.(LOC CAL ON)
CHAPTER 2 DIAGNOSIS TABLE 2-12 TROOPER COUNT SCALE - CANNOT CALIBRATE (Cont.). PROBABLE CAUSE REMEDY SYMPTOM Main PC Board is defective. If the above readings are within limits, Scale can be turned on but the Main PC Board is defective.
CHAPTER 2 DIAGNOSIS 2.7 TROOPER COUNT ERROR CODE TABLE TABLE 2-14. TROOPER COUNT SCALE ERROR CODES. ERROR CODE CAUSE REMEDY Batteries are weak. [Batt ][Low ][ Replace batteries. [Err 1][over][Load] The load on the platform ex- Remove excess load. Refer to the Trouble- ceeds the rated capacity of shooting Tables.
Page 29
CHAPTER 2 DIAGNOSIS TABLE 2-14. TROOPER COUNT SCALE ERROR CODES (Cont.). REMEDY CAUSE SYMPTOM Recalibrate scale. If the problem per- Capacity point parameter [CAP][data][Error] sists, replace the Main PC board or Load check failed during program Cell. running. Recalibrate the scale.
CHAPTER 3 REPAIR PROCEDURES 3. REPAIR PROCEDURES This section describes how to change individual components of the Trooper Scale and Trooper Count Scale. When doing this, please refer to the exploded view drawings and parts lists in section 5. Important: After replacing components, a functional check of the scale must always be carried out.
CHAPTER 3 REPAIR PROCEDURES 3.2 REPLACING THE TROOPER SCALE MAIN PC BOARD The Main PC Board is located inside the scale towards the front. To replace the Main PC Board, it is necessary to disassemble the scale. 1. Perform steps 1 through 4 of paragraph 3.1.
Page 33
+SEN JUMP 1 Figure 3-2. Trooper Scale RS Connector and Loadcell Connector Wiring. 10.Check the LFT lock switch (legal for trade) switch (S1). Make sure it is in the ON position. 11.Turn on the scale. When self testing is completed, the display may indicate an error code. This is normal.
CHAPTER 3 REPAIR PROCEDURES 3.3 REPLACING THE TROOPER SCALE LOAD CELL ASSEMBLY The Load Cell Assembly is centrally located inside the scale. See Figure 3-3. The load cell is factory assembled and the down stops are preadjusted. There are no adjustments to be made on this assembly after installation.
Page 35
CHAPTER 3 REPAIR PROCEDURES 3.3 REPLACING THE TROOPER SCALE LOAD CELL ASSEMBLY (Cont.) DC POWER JACK BATTERY BOX RS232 CONNECTOR LOAD CELL ASSEMBLY FERRITE CORE LOAD CELL LFT LOCK SWITCH RS232 CABLE BATTERY BOX LOAD CELL & CABLE POWER CONNECTOR...
CHAPTER 3 REPAIR PROCEDURES 3.4 REPLACING THE TROOPER COUNT SCALE MEMBRANE SWITCH The membrane switch is affixed to the upper housing of the scale. The scale must be disassembled to gain access to the membrane switch connections. 1. Remove power from the scale.
3.5 REPLACING THE TROOPER COUNT SCALE MAIN PC BOARD NOTE: Do not attempt this procedure unless you have the Trooper Count Scale Tools program. The Main PC Board is located inside the scale towards the front. To replace the Main PC Board, it is necessary to disassemble the scale.
Page 38
10. After the scale is turned on, the display will show Error 21 => Cal => Data. 11. Refer to Appendix B and start Trooper Count Scale Tools. 12. Select the model of the Trooper Count Scale in the Model box. 13.Enter the serial number in the serial number box.
CHAPTER 3 REPAIR PROCEDURES 3.6 REPLACING THE TROOPER COUNT SCALE DISPLAY PC BOARD The Display PC Board is located on the inside of the upper housing. To replace the Display PC Board, it is necessary to disassemble the scale. 1. Perform steps 1 through 4 of paragraph 3.4.
CHAPTER 4 TESTING 4. TESTING Before and after servicing the Trooper Scale and Trooper Count Scale, they should have an operational test and various performance tests made to ascertain whether or not the scale meets specifications. Turn the scale on and allow it warm up for at least one hour before performing these tests. Make sure the test area is free from drafts and the surface that the scale rests on is level and vibration free.
Set the scale readability to the factory default setting. This can be done by means of the Setup Menu – Reset function for the Trooper Count Scale and setting the Trooper Scale to kg. A scale passing each of these three tests complies with Ohaus specifications.
Page 43
CHAPTER 4 TESTING 4.3.1 Repeatability Test (Cont.) σ Figure 4-3. Standard Deviation ( ) Worksheet. Reading Weight (kg) Weight - Mean (Weight – Mean) SUM = SUM = Mean = SUM/10 Variance = SUM/9 σ Mean = = = = = = Variance σ...
7. Subtract the center position reading from the reading at each outer position. The difference at each test position should be within the cornerload tolerance specified in Table 4-1 for Trooper Scale and Table 4-2 for Trooper Count Scale. 4.3.3 Linearity Test The Linearity test is used to determine the linearity of the scale throughout its operating range.
480oz X 0.2oz 960oz X 0.5oz Linearity (g) +0.5g Repeatability +0.5g Off Center Load +0.5g 4.5 TROOPER COUNT SCALE SPECIFICATIONS TABLE 4-2. TROOPER COUNT SCALE SPECIFICATIONS. Model TC3RS TC6RS TC15RS TC30RS Default Capacity X Readability (lb) 6lb X 0.001lb 15lb X 0.002lb 30lb X 0.005lb...
4.6.2 Setting Trooper Count Scale Print Menu The Trooper Count Scale also contains a print menu with the following submenus: Baud Rate, Data Length, Parity Bit, Print Part No., Print Count, Print APW, Print Net, Print Tare, Print Total PCS, Print Gross, print on Stable, Print Header and Print Style.
RS232 Command Table 4-3 for Trooper Scale and Table 4-4 for Trooper Count Scale. See Figure 4-7 for connections. When the scale is connected directly to a printer, displayed data can be output at any time by simply pressing PRINT UNITS button on the Trooper Scale and Trooper Count Scale.
4. Press PRINT Unit button, the computer or a printer should record the weight value. 5. Refer to Tables 4-3 for the Trooper Scale and 4-4 for the Trooper Count Scale and try all commands and verify operation of the scale.
CHAPTER 5 PARTS LISTS 5. PARTS LISTS This section of the manual contains exploded views, and parts lists for the Trooper Scale and Trooper Count Scale. The exploded view drawings identifies the replaceable parts. NOTE: In all cases where a part is replaced, the scale must be thoroughly checked after the replacement is made.
The sample illustration indicates a 4kg mid point for a 6kg scale. (Linearity calibration for Trooper is 0, 2/3 and full capacity). Place the indicated mass on the platform. Keep the platform stable during this period.
APPENDIX A - CALIBRATION A.2 TROOPER COUNT SCALE LINEARITY CALIBRATION WEIGHT DATA COUNT The following example is with a 6kg Scale. With the Scale ON, press and hold the SAMPLE Menu button until CAL MENU is displayed. < > PT APW MR...
Page 57
APPENDIX A - CALIBRATION A.2 TROOPER COUNT SCALE LINEARITY CALIBRATION (Cont.) COUNT COUNT WEIGHT DATA COUNT < > < > < > B/G %Ac PT APW MR Place the indicated calibration mass on the pan. Press the Enter button. The sample illustration indicates a 6kg Scale using a 6kg mass.
APPENDIX A - CALIBRATION A.3 LINEARITY CALIBRATION WEIGHTS Linearity calibration weights are defined in Tables A1 and A-2 for the Trooper Scale and Trooper Count Scale. TABLE A-1 TROOPER SCALE LINEARITY WEIGHTS Cal in kg: Linearity cal (fixed) Masses Totaling TR3RS 2 &...
Scale. Through the RS232 interface, this program allows a user to configure the scale settings and library data and test output commands from a PC. Use of the Trooper Count Scale Tools is required after the main PC board of the Trooper Count Scale is replaced.
Serial No. Serial No. • Serial No. Serial No. Serial No. Up to a 6-digit serial number of the Trooper Count Scale is entered in this field. Metric Unit Setting Metric Unit Setting • Metric Unit Setting Metric Unit Setting...
Page 61
• [Read Config Data from Scale] Button [Read Config Data from Scale] Button [Read Config Data from Scale] Button Select this button to load and display the configuration data from the scale into the Trooper Count Scale Tools program. [Write Config Data to Scale] Button [Write Config Data to Scale] Button •...
APPENDIX B - TROOPER COUNT SCALE TOOLS B.3.2 Library The Library Screen offers a method to quickly and conveniently read, enter, and save individual Library records. The Library screen displays up to 60 part-files, each line representing a record that contains unique part- file information.
Page 63
APPENDIX B - TROOPER COUNT SCALE TOOLS U n i t U n i t • U n i t U n i t U n i t The selected unit of measurement. Select by scrolling through the available units: kg, lb, g or Note: Any changes to the record data will not be saved or take effect for scale operation until transferred into the scale or a file.
Page 64
Select this button to read (extract) all Library data from a file created from the [Write to File] Button in the Trooper Count Scale Tools program (refer following section). Specify the source directory, folder and file name as prompted on the following “Open” screen. Then select [Open] to execute the process.
The scale will also reflect that the operation was executed. • V (Get Version Number) V (Get Version Number) V (Get Version Number) V (Get Version Number) V (Get Version Number) Displays the model and software version of the Trooper Count Scale on the “Response from scale” field.
APPENDIX B - TROOPER COUNT SCALE TOOLS B.3.4 COM PORT The COM Port Screen allows the user to automatically set up RS232 connections, or manually specify scale print settings to establish proper communications. Automatic Setup is the default setting. The following scale Print Menu settings can be specified when...
Library Screen) when attempting to read a file that was not created from the [Write to File] Button in the Trooper Count Tools program. Open a file created from the [Write to File] Button. • Library error: Part number entry...
Page 68
APPENDIX B - TROOPER COUNT SCALE TOOLS B-10...