SCALEMATE
required weight on the center of the platform. If you
wish to change the weight to 100g, press the TARE
key before placing the weight on the scale.
4. Once the weight is confirmed, the "g" will stop
flashing on the display and calibration is complete.
IX. Troubleshooting
Problem: Scale will not turn on
Solution: Check the batteries
Problem: Displayed weight fluctuates randomly
Solution: Make sure your workspace is stable and
free of any air currents or vibration
Problem: Displayed weight is inaccurate
Solution: Perform a calibration adjustment
X. Error Codes
1. OUTZ - Zero range has shifted. Perform calibration
adjustment. If the problem persists, the load cell may
be permanently damaged due to overloading.
2. 88888 - Maximum capacity exceeded. Remove the
extra weight to avoid damaging the load cell.
3. L - Batteries are low. Replace the batteries.