Electromagnetic Immunity (IEC60601-1-2)
Immunity test
Electrostatic
discharge (ESD)
IEC 61000-4-2
Electric fast
transient/burst
IEC 61000-4-4
Surge
IEC 61000-4-5
Voltage dips, short
interruptions and
voltage variations
on power supply
input lines
IEC 61000-4-11
U
: Rated voltage of
T
test unit
Power frequency
(50/ 60 Hz)
magnetic field
IEC 61000-4-8
Note: U
is the a.c. mains voltage prior to application of the test level.
T
IEC60601-1-2
test level
±6 kV contact
±8 kV air
±2 kV for
power supply
lines
±1 kV for
input/output lines
±1 kV
differential mode
±2 kV
common mode
<5 % U
T
for 0.5 cycle
40 % U
T
for 5 cycles
70 % U
T
for 25 cycles
<5 % U
T
for 5 sec.
3 A/m (r.m.s)
Compliance
level
±6 kV contact
±8 kV air
±2 kV for
power supply
lines
±1 kV for
input/output lines
±1 kV
differential mode
±2 kV
common mode
<5 % U
T
for 0.5 cycle
40 % U
T
for 5 cycles
70 % U
T
for 25 cycles
<5 % U
T
for 5 sec.
3 A/m (r.m.s)
Electromagnetic
environment -
guidance
Floors should be wood,
concrete or ceramic tile.
If floors are covered with
synthetic material, the
relative humidity should
be at least 30 %.
Mains power quality
should be that of a
typical commercial or
hospital environment.
Mains power quality
should be that of a
typical commercial or
hospital environment.
Mains power quality
should be that of a
typical commercial or
hospital environment. If
the user of the HBP-
1100 requires continued
operation during power
mains interruptions, it is
recommended that the
HBP-1100 be powered
from an uninterruptible
power supply or
batteries.
Power frequency
magnetic fields should
be at levels
characteristic of a typical
location in a typical
commercial or hospital
environment.