Professional Blood Pressure Monitor
HBP-1300
Instruction Manual
Thank you for purchasing this OMRON Professional
Blood Pressure Monitor.
Please competely read this Instruction Manual before
using the monitor for the first time.
Read this manual to ensure the safe and accurate
use of the monitor.
For Professional Use
KEEP THIS INSTRUCTION MANUAL
NEAR THE BLOOD PRESSURE
MONITOR FOR FUTURE REFERENCE
Caution: Federal (USA) law restricts this device to sale by or on the order of a physician or licensed
healthcare practitioner

Table of Contents

Intended Use.............................................. 1
Exemptions ................................................ 1
Notes on Safety ......................................... 2
Warnings and Cautions.............................. 3
Components of the Product ..................... 10
Optional Accessories ............................... 10
Features of the Product ............................11
Features / Functions of Unit..................... 12
Connecting the AC Adapter...................... 15
Menu Mode .............................................. 16
Utility Mode .............................................. 17
Cuff Selection and Connection ................ 19
Applying the Cuff to Patient ..................... 20
Stopping the Measurement...................... 23
Displaying Last Reading .......................... 23
Managing Supplies .................................. 24
Device Maintenance ................................ 24
Accessory Care........................................ 24
Check before Use .................................... 25
Troubleshooting ....................................... 26
List of Error Codes ................................... 28
Disposal ................................................... 29
Limited Warranty...................................... 30
Factory Default Settings........................... 31
Specifications........................................... 31
Manufacturer's Declaration ...................... 33

Advertisement

Table of Contents
loading

Summary of Contents for Omron HBP-1300

  • Page 1: Table Of Contents

    Taking the Measurement in "Manual Mode" ..21 Taking the Measurement in "Auscultation Mode" ...22 Stopping the Measurement...... 23 Displaying Last Reading ......23 Thank you for purchasing this OMRON Professional Blood Pressure Monitor. Maintenance Maintenance Inspection and Safety Management ...24 Please competely read this Instruction Manual before Managing Supplies ........
  • Page 3: Introduction

    Warnings and cautions described in the instruction manual should be observed at all times. Exemptions Liability and Warranty OMRON HEALTHCARE does not accept liability and the warranty becomes void under the following circumstances: 1. Failures, damage, or injuries due to maintenance or repair work performed by persons or a repair facility not authorized by OMRON HEALTHCARE.
  • Page 4: Notes On Safety

    Notes on Safety The warning signs and symbol examples indicated below are intended to ensure safe use of the product and prevent damage and injury to you and others. The signs and symbols are explained below. Safety Symbols used in this Instruction Manual Indicates the matters in which death or severe bodily damage may arise as a result Warning of incorrect handling.
  • Page 5: Warnings And Cautions

    Do not use the cuff or AC adapter to lift the unit, it can also cause the unit to malfunction. If the unit has broken down, contact OMRON HEALTHCARE Customer Service. Do not use in combination with a hyperbaric oxygen therapy device, or in an environment where combustible gas may be generated.
  • Page 6 For the AC adapter connected to the unit, supplies, and optional devices, use only the standard accessories or OMRON-specified products. Do not use in a location with moisture, or a location where water may splash on the unit.
  • Page 7 Caution Before using the unit, verify that none of the following apply to the patient: - Poor peripheral circulation, noticeably low blood pressure, or low body temperature (there will be low blood flow to the measurement position) - The patient uses an artificial heart and lung (there will be no pulse) - An SpO sensor and the cuff are attached to the same arm - The patient has an aneurysm...
  • Page 8 Cleaning Warning When cleaning the unit, turn off the power and disconnect the AC adapter from the unit. After cleaning the unit, make sure it is completely dry before connecting to a power outlet. Do not spray, pour, or spill liquids into or onto the unit, accessories, connectors, buttons, or openings in the housing.
  • Page 9 Warnings and cautions for safe measurement Optional rechargeable battery Warning If battery fluid comes in contact with the eye, immediately flush with copious amounts of water. Do not rub. Seek medical attention immediately. Do not use the optional battery pack in any other device besides this unit. Do not throw into flame, disassemble, or heat.
  • Page 10 The accuracy of a flashing measurement value that is out of the measurement range cannot be guaranteed. Always check the patient's condition before deciding what steps to take. Do not use the cuff if it is damaged or has holes. Only OMRON GS CUFF can be used with this device.
  • Page 11 • If you sense that something is wrong with a battery, immediately move it to a safe location and contact the administrator responsible for the unit or OMRON HEALTHCARE Customer Service. • If the battery voltage is low, operation by battery may not be possible.
  • Page 12: Using The Unit Components Of The Product

    Components of the Product Before using the unit, make sure that no accessories are missing and that the unit and accessories are not damaged. If an accessory is missing or there is damage, please contact OMRON HEALTHCARE Customer Service. Main unit...
  • Page 13: Features Of The Product

    Only OMRON GS CUFF can be used with this device. Features of the Product The OMRON HBP-1300 is an affordable professional blood pressure unit that is clinically proven accurate and provides fast, reliable results and is easy to use. Key features, benefits, look •...
  • Page 14: Features / Functions Of Unit

    Features / Functions of Unit Front and back of unit Power ON / Measure blood pressure Press when the power is off to turn on the power and start blood pressure measurement. During inflation or [START/STOP] button measurement, press to stop. (Power ON/OFF) Power OFF Hold down at least 3 seconds to turn off the power.
  • Page 15 LCD Display Displays systolic blood pressure. Displays diastolic blood pressure. Pulse Displays the pulse rate. Pulse Flashes in synchronization with the pulse during synchronization icon measurement. Lights in the measurement result display and memory Irregular pulse wave display if the pulse wave interval was irregular or there icon was body movement during measurement (page 23).
  • Page 16: Installing The Optional Battery Pack

    Installing the Optional Battery Pack Warning • If battery fluid comes in contact with the eye, immediately flush with copious amounts of water. Do not rub. Seek medical attention immediately. • Do not use the optional battery pack in any other device besides this unit. Do not throw into flame, disassemble, or heat.
  • Page 17: Connecting The Ac Adapter

    Connect the AC adapter to the DC jack on the unit and the power outlet. Note: When the optional battery pack is installed If there is no problem with the following. - AC adapter - DC jack - Power outlet - Battery And the icon does not flash, contact OMRON HEALTHCARE Customer Service.
  • Page 18: System Settings

    System Settings The system settings are divided into two modes, "Menu Mode" and "Utility Mode". Menu Mode "Menu Mode" allows you to configure the "Initial Inflation Pressure Value" and "Auscultation Mode" settings. Initial Inflation Auscultation Mode Pressure Value 1. Press the [MODE] button. The "Initial Inflation Pressure Value"...
  • Page 19: Utility Mode

    1. Connect the blood pressure monitor, the standard pressure gauge, and the cuff and rubber ball. For details on the standard pressure gauge, contact OMRON HEALTHCARE Customer Service. 2. Check the blood pressure value of the blood pressure monitor and the blood pressure value of the standard pressure gauge.
  • Page 20: Non-Invasive Blood Pressure (Nibp) Measurement

    Non-Invasive Blood Pressure (NIBP) Measurement Non-Invasive Pressure Measurement Principles Oscillometric method The beat in the pulsation generated by the contraction of the heart is captured as the pressure inside the cuff to measure the blood pressure. If the cuff wrapped around the upper arm is pressurized sufficiently, the blood flow stops, but the beat of the pulsation is present and the pressure inside the cuff receives this and oscillates.
  • Page 21: Cuff Selection And Connection

    Connect the cuff tube to the NIBP connector on the unit and turn clockwise to lock. Caution Only OMRON GS CUFF can be used with this device. Note: • If the cuff tube is too short, the optional 3.5 m or 1.5 m CUFF HOSE can be connected to lengthen the cuff tube.
  • Page 22: Applying The Cuff To The Patient

    Applying the Cuff to the Patient 1. Wrap on a bare arm or over thin clothing. Wrap the cuff on a bare arm or over thin clothing. Thick clothing or a rolled up sleeve will cause inaccurate blood pressure measurements. The device can be used on either the right or left arm.
  • Page 23: Taking The Measurement In "Manual Mode

    Caution Make sure the cuff is wrapped in the correct arm position and is at the same height as the heart. A difference of 4 inches (10 cm) in height may cause a variation in the blood pressure value of up to 7 - 8 mmHg.
  • Page 24: Taking The Measurement In "Auscultation Mode

    Irregular pulse wave detection function If the pulse wave interval becomes irregular during measurement, the irregular pulse wave detection icon will light to notify you. Body movement detection function If body movement is detected during measurement, deflation stops for 5 seconds. The irregular pulse wave icon appears in the measurement result display.
  • Page 25: Stopping The Measurement

    Stopping the Measurement To stop measurement while measurement is in progress, press [START/STOP] button. Displaying Last Reading The previous measurement value (systolic blood pressure, diastolic blood pressure, and pulse rate) and whether or not an irregular pulse wave was detected can be displayed. In the case of auscultation measurement, the auscultation icon appears.
  • Page 26: Maintenance

    Maintenance Maintenance Inspection and Safety Management The HBP-1300 must be maintained to ensure functionality and to secure the safety of patients and operators. Daily checks and maintenance should be performed by the operator. (page 25) In addition, qualified personnel are necessary to maintain the performance and the safety, and to conduct periodic inspections.
  • Page 27: Check Before Use

    ■ Press each button and check that it works. Non-invasive blood pressure (NIBP) ■ Make sure that a suitable OMRON GS CUFF is attached (one that fits the circumference of the patient's arm). ■ The cuff tube is firmly connected.
  • Page 28: Troubleshooting

    An object is on top of the unit or right next to the unit. Keep the area around the unit free of objects. If the unit becomes too hot to be touched, there may be a problem in the unit. Turn off the unit power, disconnect the AC adapter, remove the battery, and contact OMRON HEALTHCARE Customer Service at 1-877-216-1336.
  • Page 29 Abnormal measurement value Cause / solution The causes below are possible. Check the patient by palpation and then repeat measurement. • Body movement (chills or other trembling) • Arrhythmia. • Noise in the cuff - A nearby person touched the patient. - Cardiac massage was being performed.
  • Page 30: List Of Error Codes

    Points to check code The cuff tube is not connected Firmly connect the cuff tube. Replace with an OMRON GS CUFF that Air is leaking from the cuff. does not leak. Did not inflate properly because the arm Have the patient not move the arm or body, or body moved during measurement.
  • Page 31: Disposal

    Medium Battery failed to charge. replace with a new battery. (page 14) Replace the battery with a new battery. If the Medium Battery voltage error error continues, contact OMRON HEALTHCARE Customer Service. Disposal Description As there is a risk of environmental pollution, follow your applicable Federal, state and local legal regulations regarding disposal or recycling of this equipment and batteries.
  • Page 32: Limited Warranty

    IMPOSED BY LAW, IF ANY, ARE LIMITED IN DURATION TO THE PERIOD OF THE ABOVE EXPRESS WARRANTY. OMRON SHALL NOT BE LIABLE FOR LOSS OF USE OR ANY OTHER SPECIAL, INCIDENTAL, CONSEQUENTIAL OR INDIRECT COSTS, EXPENSES OR DAMAGES. This warranty provides you with specific legal rights, and you may have other rights that vary by jurisdiction.
  • Page 33: Specifications

    Specifications Factory Default Settings Factory default settings and backup are as shown below. Backup : Setting is retained even if the power is interrupted. : Reverts to factory default setting if the power is turned off. Setting Pressure Value Settings Factory Setting Backup Initial Inflation Pressure Value...
  • Page 34: Power Supply

    Power supply AC adapter Input voltage range: AC 100 V to 240 V Rated Current: 1.6 A Frequency: 50/60 Hz Output voltage range: DC 6 V ±5% Optional rechargeable battery Type: 3.6 V, 1900 mAh Number of operation cycles when fully charged: 300 Measurement conditions •...
  • Page 35: Manufacturer's Declaration

    Manufacturer's Declaration The HBP-1300 is intended for use in the electromagnetic environment specified below. The customer or the user of the HBP-1300 should assure that it is used in such an environment. Electromagnetic Emissions: (IEC60601-1-2) Emission Test Compliance Electromagnetic Environment RF emission Group 1 The HBP-1300 uses RF energy only for internal functions.
  • Page 36 IEC60601-1-2 IEC60601-1-2 Electromagnetic Immunity test test level test level environment - guidance Portable and mobile RF communications equipment should be used no closer to any part of the HBP-1300, including cables, than the recommended separation distance calculated from the equation applicable to the frequency of the transmitter.
  • Page 37 Note: POTENTIAL FOR RADIO / TELEVISION INTERFERENCE (for U.S.A. only) This product has been tested and found to comply with the limits for a Class B digital device, pursuant to part 15 of the FCC rules. These limits are designed to provide reasonable protection against harmful interference in a residential installation.
  • Page 38 Recommended Separation Distances: Recommended separation distance between portable and mobile RF communications equipment and the HBP-1300 The HBP-1300 is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The customer or the user of the HBP-1300 can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF communications equipment (transmitters) and the HBP-1300 as recommended below, according to the maximum output power of the communications equipment.
  • Page 40 For Customer Service Call Toll Free 1-877-216-1336 Manufactured by OMRON HEALTHCARE Co., Ltd. OMRON HEALTHCARE, INC. Distributed by 1925 West Field Court, Lake Forest, IL 60045, USA Made in China. © 2012 OMRON HEALTHCARE, INC.