Note:
Setup
• Read and understand the manual for each optional accessory. This
manual does not contain cautionary information for optional accessory.
• Exercise caution with the cables and arrange so that the patient does not
become entangled or bound.
Before use / during use
• Check the following after turning on the power:
- No smoke, abnormal odor, or abnormal sound is emitted.
- Press each button and verify that it operates.
- For functions that cause icons to light or flash, verify that the icons light
or flash.
- Measurement can be performed normally, and measurement error is
within the tolerance value.
• If the screen is not displayed normally, do not use the unit.
• When recycling or disposing of parts (including batteries) of the unit,
follow local government rules and regulations.
Cleaning
• For cleaning, see page 23.
Non-Invasive Blood Pressure (NIBP) measurement
• If the patient has acute inflammation, a pyogenic ailment, or an external
wound at the location where the cuff is to be wrapped, follow the
instructions of the doctor.
• Non-Invasive Blood Pressure measurement (NIBP) is performed by
compressing the upper arm.
Some people may experience intense pain, or transient spotting caused
by subcutaneous hemorrhaging (bruising) may appear. The spotting will
disappear with time; however, it may be appropriate to inform patients for
whom this may be a concern that spotting sometimes occurs, and if
necessary, refrain from measurement.
• To measure correctly, it is recommended that the patient relax and not
talk during measurement.
• To measure correctly, it is recommended that the patient rest quietly for 5
minutes before measurement.
9