Advertisement

Error
Symptom
No.

Actual measurement
time is different from
the displayed
measurement time.
2
Measurement value on
the LCD disappears.
3
During USB
communication:
Cannot download data
output from this
instrument on the PC.
Cannot enter
commands or data from
the PC to this
instrument.
4
Same error message
appears repeatedly.
Item to Check
Has the power source
been supplied securely?
Did you interrupt the
measurement?
Has the USB cable
been connected
securely?
Has the USB cable
been disconnected?
Has remote mode been
canceled?
Has the program been
prepared correctly?
Check the appropriate
corrective action for the
error message.
84
Corrective Action
The displayed measurement
time is the remaining time. The
actual measurement time may
be different from the displayed
time depending on the mode
setting of measurement time.
Connect to a stable power
source and insert the AC adapter
plug securely.
When starting continuous
measurement, press the
measurement button securely.
Do not press ESC key.
Connect this instrument and the
PC securely.
Replace the USB cable.
Send the connection command
from the PC to this instrument
and switch to remote mode. Use
the data management software
CS-S0w in the standard
accessories.
Refer to communication
specifications and check the
program. Use the data
management software CS-S0w
in the standard accessories.
If the symptoms are not
remedied, please contact the
nearest KONICA MINOLTA
authorized service facility.