Error Messages - Omron HCG-801 Instruction Manual

Portable, cordless, single-channel ecg monitor
Hide thumbs Also See for HCG-801:
Table of Contents

Advertisement

HCG-801-E.book Page 48 Tuesday, December 13, 2011 5:35 PM
10. Care and Maintenance
Message
SD is write-protected.
No saved data.
Set clock.
No saved data.
Set clock.

Error Messages

Error #
Message
Please insert new
01
batteries.
Device memory error
11
Cannot read data.
Device memory error
12
Cannot erase data.
Device memory error
13
Cannot write data.
48
Cause
Refer to the instruction supplied
The unit is unable to access data
with the memory card and insert
because the SD memory card
the SD memory card with write-
inserted is write-protected.
protection turned off.
Insert new batteries and reset the
The battery power has run low
Date and Time and other
and the ECG data stored on the
settings. Any ECG data that was
unit has disappeared.
stored on the unit cannot be
recovered.
The battery power has run low
Insert new batteries and reset the
and the Date and Time setting on
Date and Time from the SET
the unit has disappeared.
menu.
The battery power has run low
Insert new batteries and reset the
and the ECG data stored on the
Date and Time and other
unit has disappeared or the clock
settings.
has been reset.
Cause
Replace both batteries with new
The batteries are worn out.
ones.
An error occurred while
trying to read ECG data
stored in unit memory.
Try again.
An error occurred while
If the error still occurs, consult
trying to erase ECG data
your OMRON service
stored in unit memory.
representative.
An error occurred while
trying to save ECG data to
unit memory.
Remedy
Remedy