Other Location Considerations - Kodak Dryview 5800 User Manual

Laser imager
Hide thumbs Also See for Dryview 5800:
Table of Contents

Advertisement

Specifications
Other Location
Considerations
5-2
Additional factors that influence where the imager is located include
proximity to patients, flammable materials, liquids, and other equipment.
CAUTION
:
This equipment is not contained in a sealed cabinet. Do not use this
equipment in locations where it can come in contact with liquids,
including body fluids.
IMPORTANT:
This equipment is intended to connect to other medical devices. Only
an authorized service provider may install this equipment.
Do not locate the imager within 2.0 m (6.6 ft) of where a cell phone
will be used, even if the cell phone is separated from the equipment
by a wall.
Do not locate the imager within 4.0 m (13.1 ft) of a microwave oven.
Electromagnetic radiation from a microwave oven is only an issue if,
after the oven door is closed and latched, the seal does not maintain
an electromagnetic tight fit between the oven door and the oven main
housing. Determining if the seal has an electromagnetic tight fit
requires special detection equipment.
Do not locate the imager in the presence of flammable anesthetics,
oxygen, or nitrous oxide. The imager does not have a gas-sealed
electronics enclosure and could ignite any flammable or explosive
gases present in the imager's environment.
Do not locate the imager in close contact with MRI devices due to
possible high magnetic fields near an MRI unit. The magnetic field in
the area where the imager is installed must be less than 50 Gauss.
1.83 m (6.00 ft)
Do not locate the imager closer than 1.83 m (6.00 ft) from a patient
bed or chair.
Do not locate the imager in a room that is in direct sunlight.
Do not substitute or modify any part of the imager.
2G0733
2011-09-09