Environmental Requirements

Temperature

Relative Humidity

Altitude

Surface Condition

Environmental Effects

Laser Specifications

2011-09-09
Operating: 15–33 °C (59–91 °F)
Storage: From –40 through 60 °C (From –40 through 140 °F)
Operating: 20–80 % RH, noncondensing
Storage: 10–90 % RH, noncondensing
30 m (100 ft) below sea level to 2438 m (8000 ft) above sea level
The surface must be a level (must be level within 1 °) table top or counter
capable of supporting the weight of the imager. For the imager to operate
with no image degradation, surface vibration levels cannot exceed the
following:
2
Freq. Range
G
/Hz
-7
1–50 Hz
1 x 10
-5
50–200 Hz
6 x 10
-6
200–650 Hz
5 x 10
-6
650–1000 Hz
1 x 10
Heat Dissipation: total heat dissipation for the imager is
883 BTU/hour
Acoustical Noise:
– Less than or equal to 75 dB-A at 1 m during a sound burst of 1 sec
or less
– Less than or equal to 65 dB-A at 1 m during normal operation
– Less than or equal to 50 dB-A at 1 m during idle mode
Wavelength : Minimum 800 nm, nominal 810 nm, maximum 820 nm
Laser Beam Divergence: Maximum 32 degrees, minimum 6.8 degrees
Laser Maximum Power: 120 mW
Radiation class: Class 3B invisible laser radiation
Type: Diode
2G0733
Direction
Vertical or horizontal