AWS SG-300 User Manual page 5

(300g x 0.1g) spoon scale

Advertisement

VII.
Calibration
Initial calibration is performed at the factory. Calibration
should only be performed by experienced users and only
if the scales readings are off on a known weight such as
a calibration weight. Please note that accuracy will also
depend highly on the angle that you hold the scale. It should
always remain level:
1. Place the required calibration weight on the scale
and note the displayed value.
2. If the displayed value is not within acceptable
tolerance (see chart on next page), adjustment is
required.
Model
SG-300
VIII.
Adjustment
1. Press and hold the MODE button to enter calibration
mode.
2. The display will flash "CAL". Make sure the spoon is
empty and press MODE again.
3. The display will flash "200.0g". Place a 200g
calibration weight in the weighing spoon and keep
the scale level and stable.
4. The display will flash "PASS" then return to normal
weighing. Calibration is complete!
Acceptable
Tolerance (±)
0.3g
Calibration
Weight
200g
Recommended
Accuracy Class
OIML M2