m
warnInG
Emergency operation must only be used in emergencies, and then with great care. When the seat
is operated in the emergency mode, all safety functions in the control system are bypassed. There
is a risk of crushing and a risk of damage to the wheelchair.
Never drive the wheelchair and use the emergency operation cable at the same time.
7
Use the control lever on the emer-
gency operation cable to operate
the function you chose to connect
to one of the holes, see picture 7.
8. If required, change to a new func-
tion by first disconnecting the
emergency operation cable and
then reconnecting it as shown
before to the appropriate control
module. Then repeat steps 5-7.
Disconnecting the
emergency operation
cable
1. Disconnect the end of the emer-
gency operation cable from the
connection to the adjustment
device cable.
2. Disconnect the other end of the
emergency operation cable from
the connection to the charging
socket on the control panel.
3. Connect all adjustment device
cables to their respective control
modules.
4. Replace the back cover of the
wheelchair with two Allen screws.
5. Replace the bumper with two Allen
screws.

Emergency operation of seat functions

Maneuvering with the control lever
of the emergency operation cable