USP EM-900 WA User Manual page 10

Wanderer alert, 40 patient monitor station

Advertisement

01 : JOE SMITH
Then you are given a choice of
TURN OFF?
YES(1) NO(2)
Press '1' to TURN STATION OFF
Press '2' to leave STATION ON
(Press '2' No, to leave STATION ON)
VIEW DATA LOG – Examining ALARM log, Tx MISSING log, and MISSED codes
This data logging feature allows you to evaluate important information collected over time by storing:
The total number of times each transmitter has been tripped. The maximum count is 99. Beyond 99 the
display will show ">". You may reset the count to 0 at any time.
The total number of times a transmitter is found to be missing when the receiver fails to record 30 check-in
codes (in a row) sent at 1 minute intervals, causing the Tx MISSING counter to be incremented by one for
each such occurrence. The maximum count is 99, which may be reset at any time.
The highest number of missed check-in codes in a row of less than 30 recorded during the monitoring
process. This is an important measure by which you can judge how reliable the wireless link is for individual
transmitter Stations. Remember, a Tx MISSING alarm will be given if there are 30 missing check-in codes in a
row. The maximum count is 30. Beyond 30 the display will show ">". You may reset the count to 0 at any time.
As a rule of thumb, the largest number of missed codes for each transmitter should not exceed 15 over a 24-
hour period. If it does, you may have to reduce the distance or change location of that particular transmitter to
minimize the risk of false alarming.
Note: Data logging only applies to receiving channels that are enabled.
Screen will then display
1 MON/NEW ENTRY
2 ON-OFF/VU LOG
Press '1' to MONITOR or ENTER NEW PROGRAM
Press '2' to TURN STATION ON – OFF or VIEW STATION LOG
(Press '2' to TURN STATION ON – OFF or VIEW STATION LOG)
Screen will then display
ENTER STATION #:
(01 - 40)
Enter 01 (for station 01).
STATION: 01
Screen will then display
1 STATION ON - OFF
2 VIEW DATA LOG
Press '1' to CHANGE STAION CONFIGURATION from ON to OFF or OFF to ON
10