Transport and storage conditions:

To prevent damage!
The Battery Drill System 22/4400-00 and attachments must be transported in such a way that no damage can occur. This particularly means en-
suring suitable packaging. The same attention must be paid to correct storage. This device must only be kept in a dry, closed place!
Safety instruction: Remove the battery from the Battery Drill System
1. Before sterilization
2. If any fault occurs
3. For transport or storage when not in use
4. Before sending for servicing
If, for any reason, the drill is stored or transported differently, this may result in damage which is not covered by the manufacturer's
or dealer's warranty.

Troubleshooting:

Fault
Motor runs too slowly or torque
is to low
Lower trigger not pressed
Motor does not run
Drill cannot be switched from
reverse to clockwise rotation
Repairs:
LEIBINGER GmbH&Co.KG only accepts responsibility for the safety, reliability and performance of the Battery Drill System 22/4400-00 if
the Battery Drill System 22/4400-00 is used in compliance with the instructions in the User's Manual.
installation, additions, adjustments, modifications and repairs are carried out by LEIBINGER GmbH&Co.KG or by qualified persons ex-
pressly authorized by LEIBINGER GmbH&Co.KG.
Only original spare parts from LEIBINGER GmbH&Co.KG may be used for repairs.
The operator is required to obtain a list detailing the type and scope of work performed, with the company stamp, date and signature
of the repairer.
SERVICING:
We recommend sending the Battery Drill System 22/4400-00 to LEIBINGER GmbH&Co.KG for servicing after a year or after 120 sterilization cycles
(surgical operations).
WARNING!
All equipment must be cleaned and sterilized before sending for servicing!
4
Cause
NiMH battery is to weak
NiMH battery is defective or
depleted
NiMH = nickel-metal hydride
Motor defective
Control system defective
Trigger switch defective
Gears or friction bearing
defective
Solenoid switch (trigger) defec-
tive
Control system defective
Symptom
No
sound
when
trigger
is
pressed
(R or L trigger pressed)
Remark:
The lower trigger has a double
function: main switch and re- verse
rotation( upper trigger = clockwise
rotation)
Reversing function does not
react
Continuous running or wrong
direction of rotation
Rectification
Recharge Battery 22/4415-00;
22/4416-00
Inspect Charger 22/4413-00,
22/4414-00 or 22/4411-00,
22/4412-00 If necessary, change
the battery
Send to manufacturer for re-
pair
Send to manufacturer for re-
pair