Maintenance - Leibinger 22/4400-00 User Manual

Battery drill system
Table of Contents

Advertisement

Cleaning, disinfection, sterilisation:
Cleaning and disinfection
Wipe the surfaces immediately after use with a lint-free cloth moistened with a cleaning agent or disinfectant, or use a disinfectant
spray, to prevent exudates or other contamination from drying on. For this manual wipe--down, use a cleaning agent or disinfectant
approved for surface disinfection under national regulations (DGHM - German Association of Hygiene and Microbiology). Please
follow the manufacturer's instructions for correct dilution, exposure time etc. In the case of heavy contamination, the Battery Drill System
22/4400-00 can also be cleaned under running water (machine cleaning is not permitted).
Note:
Excessive exposure to saline solutions or solutions containing iodine or chloride may cause harmful reactions. Damage may also be
caused by strong acids and alkaline solutions or incorrectly used disinfectants. The quality of the water used for cleaning and rinsing
is another possible factor affecting corrosion or surface damage for the drill and attachments. The Battery Drill System 22/4400-00 must never be
immersed in water or cleaning solutions, or boiled. Always close the cap of the Battery Drill System 22/4400-00
before cleaning!
CAUTION!
Machine cleaning and ultrasonic cleaning in thermo disinfectors are not permitted!
Disinfection is required after every surgical intervention to destroy microorganisms. Today, health professionals and patients are at
constantly increasing risk, so appropriate cleaning and disinfection of these drills and attachments is extremely important. If in doubt,
contact LEIBINGER GmbH&Co.KG and the manufacturer of the disinfectant and cleaning agent used!
Note:
Please observe the "Instructions for use of attachments" (with the exception of the Rechargeable Battery 22/4415-00; 22/4416-00 and Desktop
Charger 22/4413-00, 22/4414-00 or 22/4411-00, 22/4412-00). See the section "Cleaning, disinfection and sterilization of attachments".
Sterilisation:
The Battery Drill System 22/4400-00 with end cap and funnel can be sterilized at up to 2 bar at 134°C in the steam sterilizer. It is advisable to
place the equipment on a cloth or wrap it in a cloth to prevent external or internal damage. To ensure proper sterilization, the end cap
must be screwed onto the Battery Drill System 4-807.
CAUTION!
The Rechargeable Battery 22/4416-00 may never be sterilized because if it is heated to above 65°C, it will be irreparably damaged.

Maintenance:

The Battery Drill System 22/4400-00 itself requires no maintenance. Nevertheless, we recommend annual inspection. This must only be entrusted
to the manufacturer or qualified persons authorized by the manufacturer. The maintenance interval depends on the number of times
the Battery Drill System 22/4400-00 is sterilized.
Note:
The battery compartment of the Battery Drill System 22/4400-00 has a watertight cap that also keeps out germs. You should, therefore, regularly
inspect the sealing ring on the cap and the sealing surface on the handle.
3

Advertisement

Table of Contents
loading

Table of Contents