Battery drill system

Advertisement

Quick Links

Date: 10.10.2011 
BATTERY DRILL SYSTEM
1253
User Manual
Easy to use
Ergonomic design
Simple, compact battery packs with high capacity
Small, robust drill with an extensive range of attachments
Steam-sterilisable

Advertisement

Table of Contents
loading

Summary of Contents for Leibinger 22/4400-00

  • Page 1 Date: 10.10.2011  BATTERY DRILL SYSTEM 1253 User Manual Easy to use Ergonomic design Simple, compact battery packs with high capacity Small, robust drill with an extensive range of attachments Steam-sterilisable...
  • Page 2: Intended Use

    Intended use: The Drill 22/4400-00 is used in bone surgery for drilling and tapping, inserting and removing screws and placing Kirschner wires. The drill has a cannulated drive shaft, so that drill wires can be briefly inserted and fed through the drill (diameter of cannulation: 3.2mm) for safe working.
  • Page 3: Maintenance

    Sterilisation: The Battery Drill System 22/4400-00 with end cap and funnel can be sterilized at up to 2 bar at 134°C in the steam sterilizer. It is advisable to place the equipment on a cloth or wrap it in a cloth to prevent external or internal damage. To ensure proper sterilization, the end cap must be screwed onto the Battery Drill System 4-807.
  • Page 4: Transport And Storage Conditions

    To prevent damage! The Battery Drill System 22/4400-00 and attachments must be transported in such a way that no damage can occur. This particularly means en- suring suitable packaging. The same attention must be paid to correct storage. This device must only be kept in a dry, closed place! Safety instruction: Remove the battery from the Battery Drill System 1.
  • Page 5: Technical Data

    It is hereby certified that the Battery Drill System 22/4400-00; 22/4410-00 has been tested and is interference-suppressed in compliance with the EMC provisions of EN60601-1-2. This Battery Drill System 22/4400-00; 22/4410-00 has been tested by our notified body for a Class II.a product with internal or external energy source, as per 93/42 EEC Annex II.3...
  • Page 6: Key To The Symbols

    Disposal: LEIBINGER GmbH&Co.KG will dispose of the Battery Drill System 22/4400-00 and attachments (without any environmental contamination) in accordance with the provisions of MED. UMW LEIBINGER GmbH&Co.KG will deal with disposal! Key to the symbols: Symbol 1 Observe instructions in User´s Manual!
  • Page 7 OVERVIEW: 6. Clockwise rotation 7. Main switch and reverse rotation 1. Battery Drill 2. Battery Compartment 3. Rechargeable Battery 4. End cap 5. Chuck with Drill Bit 6. Clockwise Rotation 7. Main switch and reverse Rotation Important To obtain full power from the drill during the surgical operation, always use freshly charged batteries. Preparation Step 1.
  • Page 8 Overview of attachments: NOTE: All the clutches and chucks for the Battery Drill System are mounted in the same way! • Step 1. Slide the clutch collar (1) in the direction indicated by the arrow • Slide the chuck (2) onto the shaft (3) as far as it will go •...
  • Page 9 2.) 22/4416-00 Standard rechargeable battery 3.) 22/4417-00 Funnel for inserting Rechargeable Battery 22/4415-00 or 22/4416-00 into Battery Drill System 22/4400-00 (to avoid contamination of the Battery Drill System. If it is not used, there may be a loss of sterility) 4.) 22/4413-00...
  • Page 10: Care And Maintenance

    Cleaning, disinfection and sterilization of attachments: (except Rechargeable Battery 22/4415-00; 22/4416-00 and Desktop Charger 22/4413-00, 22/4414-00 or 22/4411-00, 22/4412-00 because these products are not categorized as medical devices ) CLEANING AND DISINFECTION: CARE AND MAINTENANCE Care and maintenance are vital to keep the attachments in good condition and fully functional! Causes of corrosion with stainless steel used for instruments.
  • Page 11 1. MECHANICAL CLEANING: Mechanical cleaning is the preferred method. Strictly follow the instructions supplied by the manufacturer of the cleaning machine. Prepare the instruments and attachments for mechanical cleaning. Multipart instruments and attachments that can be dismantled must be taken apart or opened for cleaning. All tips and delicate parts should be cleaned manually. Do not overfill wire trays; place heavy parts on the bottom of the tray.
  • Page 12: Fault Rectification

    Only store in the original packaging, in a dry, closed room. Never sterilise the BATTERY 4-809/4-810 or the Charger 22/4413-00, 22/4414-00 or 22/4411-00, 22/4412-00. Only use compatible BATTERIES 22/4415-00; 22/4416-00 from LEIBINGER GmbH&Co.KG. Combination with other makes is forbidden. Do not immerse Charger 22/4413-00, 22/4414-00 or 22/4411-00, 22/4412-00 or...
  • Page 13: Safety Instruction

    (3) of the Charger 22/4413-00, 22/4414-00 or 22/4411-00, 22/4412-00 against moisture. Never immerse in liquid or clean under flowing water! Repair: LEIBINGER GmbH&Co.KG only accepts responsibility for the safety, reliability and performance of the Charger 22/4413-00, 22/4414-00 or 22/4411-00, 22/4412-00 and BATTERY 22/4415-00; 22/4416-00 if: •...
  • Page 14 Technical Data: Type of equipment: Chargers 22/4413-00, 22/4414-00 or 22/4411-00, 22/4412-00 Rated voltage 240 V and 115 V Rated frequency 50 HZ 14 VA Rated input Rated output voltage 12 V Rated output current 1100 mA max. Protection class IPX0 Degree of protection, as per DIN EN 529 Function The Chargers 22/4413-00, 22/4414-00 or 22/4411-00, 22/4412-00 are controlled and monitored by a microcontroller.
  • Page 15 Notice:...