Troubleshooting - Barton RCL-4002 Operation Manual

Ready care lift
Table of Contents

Advertisement

1. Ready Care Lift will not operate in either direction:
-ensure the actuator (motor) is plugged in and secure
-emergency switch is turned off (turn button clockwise)
-check battery level indicator
-ensure the hand control is connected to the electronics pack
-check for cuts in the hand control wire
2. Ready Care Lift operates in "UP" position only:
-small switch at the base of the actuator is engaged. To disengage, lift up on
actuator and lift arm. This switch should ride on the bolt which secures the bottom of
the actuator to the lift
-hand control is damaged. Check 'UP' button for damage
3. Ready Care Lift does not lower by hand control but "EMERGENCY LOWERING" is
operable.
-hand control is damaged
-emergency stop is engaged
4. Ready Care Lift will not raise with patient but will lower.
- battery needs charging, check battery level indicator, charge battery
- patient is heavier than capacity of lifter
5. Ready Care Lift operates but very slowly.
- battery needs charging, check battery level indicator, charge battery
6. Ready Care Lift pulls to one side while pushing.
- check all castors for free movement, remove debris which may impede movement.
- check castors for damage such (bent, loose components)

TROUBLESHOOTING