Spare Parts Listing
Scale is not transmitting data to the host
device
1. Check cable connection at both the rear
of the scale and the host device.
2. Check communication setting and baud rate
on both scale and software.
3. Perform I/O loopback test.
4. Replace the cable.
5. Replace the main board.
The ZERO key and the TEST key do not
function
1. Open display enclosure and verify that the
keypad cable is still installed correctly.
2. Replace the display panel.
3. Replace the display PCB.
4. Replace the display cable.
5. Replace the main PCB.
DESCRIPTION
Keyboard Panel
Display PCB
RS-232 PCB
4 Bit Parallel PCB
40 Bit Serial PCB
Power Supply - 6702/6710 1148-15535
Power Supply - 6720
RS-232 Cable - PC
6700-7 Kg Loadcell
6700-15 Kg Loadcell
6700-30 Kg Loadcell
6700-60 Kg Loadcell
Remote Display Kit
Remote Pole Kit
Remote Pole/Displ Kit 7'
Remote Pole/Displ Kit 14' 7300-16864-02
6702/6710 Fr/Rear Mt. Bkt.7200-14829
6700 Feet
6702/6710 Frt Mt. Bkt.
Model 6700 Family User's Manual
PART NUMBER
1163-13205
7405-15465
7405-14704-2
7405-14653-2
7405-14864-2
1148-15536
1140-13842
7154-16323-07
7154-16323-15
7154-16323-30
7154-16323-50
7300-16577-01
7200-14837
7300-16864-01
7075-10256
1062-17061
27