Power supplies
Table of Contents

Advertisement

Specifications
Safety Features
No load detection:
Sudden load change
detection:
Overload/short
circuit protection:
Ground Leak Detection:
(PowerPac 200 only)
Auto power up
after power failure:
Safety
Compliance:
EMI
Display Functions:
Function Modes:
Environmental
Operating Temp.:
Humidity:
Dimensions:
Weight:
14
Indicated by error message on LED display
Indicated by error message on LED display
Indicated by error message on LED display; fuse on both hot and neutral
Indicated by error message on LED display if the ground current is greater
than 500 µA.
User-selectable, setup values maintained
I.E.C. 1010
Conforms to CE standards for Emissions and Immunity;
tested only at 220 V. See Declaration of Conformity for details.
TUV EMC certification.
3-digit LED displays voltage, current, time
Constant voltage, constant current, timer
User-selectable automatic power-up after power failure
0 - 40°C
0 - 95%, in the absence of condensation
29 (L) x 21 (W) x 8 (H) cm.
Unit is stackable
PowerPac 200: 2.4 kg
PowerPac 300: 1.8 kg