Sensor 1
Type
Sensor 1 ID
Sensor 2
Type
Sensor 2 ID
RTD 1 Type
Sensor
Setup
RTD 1 Lead
Wire
Sensor 1
Bias
RTD 2 Type
RTD 2 Lead
Wire
Sensor 2
Bias
<Return>
Year
Month
Day
Dev Install
Date
Install Date
Write Date
Page 32
mV, TC, RTD,
Ohm
Sensor ID for Input
1
mV, TC, RTD,
Ohm
Sensor ID for Input
2
2-Wire, 3-Wire, 4-
Wire
Resistance of Input
1 RTD lead wires
Bias value which is
applied to the Input
1 measured value
2-Wire, 3-Wire
Resistance of Input
2 RTD lead wires
Bias value which is
applied to the Input
2 measured value
Enter the current year.
# # # #
This item will only be visible if no Install
Date has been written to the transmitter.
Select the current month.
January thru
This item will only be visible if no Install
December
Date has been written to the transmitter.
Enter the day of the month.
# #
This item will only be visible if no Install
Date has been written to the transmitter.
If no Install Date has been set in the
transmitter, this value is a preview of
the Year, Month, and Day entered
dd-mm-yyyy
above. Otherwise, this is the Install
Date that was previously written to the
transmitter.
Press ENTER to write the Install Date to the transmitter.
CAUTION: The Install Date can only be written once in
the life of the transmitter. You cannot erase or overwrite
the Install Date once it has been written.
STT850 Temperature Transmitter User's Manual
User entered value for the
resistance of each leg. Used only
for 2-wire RTD or 2-wire ohm input
types.
Per the Engineering Units selected
for this input.
User entered value for the
resistance of each leg. Used only
for 2-wire RTD or 2-wire ohm input
types.
Per the Engineering Units selected
for this input.
Revision 1

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents