Changing the control unit
5.
Fasten the cables under the chair and
around the base of the seat lift using cable
ties.
NB! The radius of the curve of the cable
must be at least 80 mm when the seat lift is
in its highest position.
6.
Connect the connections on the right side (2
contacts). See the marking on the cables.
NB! Ensure that the contacts go all the way
in.
7.
Connect the connections on the left side (3
contacts). See the marking on the cables..
NB! Ensure that the contacts go all the way
in.
8.
Reconnect the contacts to the battery.
9.
Replace the chassis cover.
Main fuse
Red cable from
main fuse
Black cable = Bat -
Black cable - Bat -
Rear
28
80 mm
Black cable from main fuse
+
Red cable - Bat +
Front
+
Red cable - Bat +

Hide quick links: