4.3. EMC Information

Medical electrical equipment should be used with precautions according to EMC,
and must be installed according to the EMC notices disclosed in this manual as
mobile RF transceivers could adversely affect it.
Directive and declaration of manufacturer – Electromagnetic Emission
The OMRON M24/7 is suitable for use in the speci¿ ed electromagnetic environment. The
purchaser or user of the OMRON M24/7 should assure that it is used in an electromagnetic
environment as described below
Emission test
Radiated and conducted
RF emission
CISPR 11
Radiated and conducted
RF emission
CISPR 11
Harmonic emission
IEC61000-3-2
Voltage À uctuations /
Flickers
IEC61000-3-3
22
Compliance
Electromagnetic Environment
Group 1
OMRON M24/7 uses RF energy only for its
internal function. Therefore, the emission
is very low and not likely to cause any
interference in nearby electronic equipment.
Class B
OMRON M24/7 is suitable for use in domestic
establishments and in establishments directly
connected to the low voltage power supply
network which supplies buildings used for
domestic purposes.
Not applicable
---
Not applicable
---

Hide quick links: