Overhead Luminaire Installation
The following describes single- and dual-overhead luminaire installation.
Do not install the overhead luminaire under a fire sprinkler system. Dissipated heat from the
luminaire may cause the sprinkler system to activate.
1. Carefully unpack and remove the SpectraLight III from its shipping container.
2. Remove the eyebolts (four for the single overhead, eight for the dual overhead) from the accessory package and
screw them into the weld nuts located on top of the unit. Refer to Figure 9.
In the following step, ensure adequate support. Each unit weighs 92 lbs. (41.7 kg).
3. Attach chains (not supplied) to each of the eyebolts and suspend the unit(s) at the appropriate height and angle
over the work area. The luminaire should be installed so that it is a minimum of two feet from any wall. Refer to
Figure 10.
30° ± 3°
30° ± 3°
Figure 9. Overhead Luminaire
31" ± 2"
6'0" ± 2"
Table
Figure 10. Recommended Overhead Luminaire Installation