4. Enter the scale program mode. Power down the unit and slide the scale program switch to the program
position. If the system is a dual cable system, disconnect the host to scale RS232 cable from the unit.
5. Reinstall the platter and power up the unit.
6. Enter bar code calibration mode. Use the vertical window to scan the following bar code. The scanner will
beep once as it enters the bar code calibration mode and the calibration utility will start.
Bar Code
Calibration Mode
³ 9 9 6 0 0 0
Note: If a razz tone sounds, an error has occurred. Refer to Diagnostic Indicator Display; Error Codes
starting on page 36 for additional information.
See Power Source caution statement located on page 12 of this manual.
Figure 50. Entering the Scale Program Mode
Figure 51. Platter Installation and Power Up
Figure 52. Entering Bar Code Calibration Mode