Intended use
The cleaning result must ensure that
the instruments have been disinfected
correctly, that no residues are left
behind, that subsequent sterilisation
can be carried out and that the
instruments can be used again safely.
Medical products are best reprocessed
using the DESIN varioTD programme.
The use of a suitable carrier (mobile
unit, module, insert etc.) is important to
ensure the adequate cleaning of
instruments and products. Examples
are given in the section "Areas of
application".
The washer-disinfector is programmed
to carry out the final rinse with mains
water or with processed water of a
quality to suit the application (e.g. aqua
destillata, purified water, de-ionised
water, demineralised water). The water
quality is of particular importance for
applications requiring analytically clean
laboratory glassware.
This machine complies with EN ISO
15883 for validation purposes.