Plumbing

Connection to the water inlet

The water in the machine must
not be used as drinking water.
– The machine must be connected to
the water supply in strict accordance
with current local and national water
authority regulations.
– If the iron content in the water is too
high, there is a risk of corrosion
occuring on items being processed
in the machine, as well as in the
machine itself.
If the chloride content of the water
exceeds 100mg/l, the risk of
corrosion to items being processed
in the machine will be further
increased.
– If there are too many impurities in the
local water supply, it will be
necessary to operate the steam
condenser with demineralised water.
– This machine is constructed to
comply with German water
regulations (DVGW), and may be
connected to a suitable supply
without an extra non-return valve if
national regulations permit.
– The minimum flow pressure for a
cold and hot water connection is
100 kPa (1.0 bar) and with an AD
water connection it is 60 kPa (0.6
bar).
74
– The recommended flow pressure
for a cold and hot water connection
is
250 kPa (2.5 bar) and for an AD
water connection is
bar). Otherwise fill times will be too
long.
– The maximum permitted static
water pressure is 1000 kPa (10 bar).
– If the water pressure is not in the
range specified above (the
indicator light may come on and the
fault code F E will appear in the
display) please contact Miele for
advice.
– The machine is supplied as standard
for connection to a cold (coded blue)
and a hot (coded red) water supply
up to a max. temperature of 70 °C.
Connect the inlet hoses to the
stopcocks for the hot and cold water
supplies.
– If a hot water supply is not available,
the inlet hose coded red should also
be connected to the cold water
stopcock.
– The inlet hose for the steam
condensor (without non-return valve)
must only be connected to the cold
water stopcock.
– AD water connection - see the
following page.
– Stopcocks with 3/4 BSP male thread
are to be provided on site. They
should be easily accessible so that
the water supply can be turned off
when the machine is not in use.
150 kPa (1.5
/