Figure 4-1
Table 4-2
/
/
-
-
/
Low Voltage Power Supply Voltages (on the bottom of the oscilloscope)
Power Supply Voltage Tolerances
Supply Voltage
Tolerance
+5.1 V
±250 mV (4.85 V to 5.35 V)
-5.2 V
±156 mV (-5.04 V to -5.36 V)
+15.75 V
+1.260 V, -787 mV (+14.963 V to +17.010 V)
+3.4 V
±100 mV (+3.30 V to +3.50 V)
+8.2 V
±82 mV (+8.12 V to +8.28 V)
measure at Printer Power connector on rear panel
of instrument
-
/
-
-
E
Calibrating and Adjusting
To adjust the power supply