Agilent Technologies 54641A Service Manual page 16

Oscilloscopes and mixed-signal oscilloscopes
Hide thumbs Also See for 54641A:
Table of Contents

Advertisement

* Denotes Warranted Specifications, all others are typical. Specifications are valid after a 30-minute warm-up period and
±10 °C from firmware calibration temperature.
Vertical System: Analog Channels
Analog channels
Bandwidth (-3dB)*
ac coupled
Calculated rise time
(= 0.35/bandwidth)
Single Shot Bandwidth
1
Range
Maximum Input
Offset Range
Dynamic Range
Input Resistance
Input Capacitance
Coupling
BW Limit
Channel-to-Channel Isolation
(with channels at same V/div)
Probes
Probe ID (Agilent/HP &
Tek Compatible)
1
2 mV/div is a magnification of 4 mV/div setting. For vertical accuracy calculations, use full scale of 32 mV for 2 mV/div sensitivity setting.
General Information
Vertical System: Analog Channels
Ch1 and 2 simultaneous acquisition
54641A/41D: dc to 350 MHz
54642A/42D: dc to 500 MHz
54641A/41D: 3.5 Hz to 350 MHz
54642A/42D: 3.5 Hz to 500 MHz
54641A/41D: ~1.0 ns
54642A/42D: ~700 ps
54641A/41D: 350 MHz maximum
54642A/42D: 500 MHz maximum
2 mV/div to 5 V/div
CAT I 300 Vrms, 400 Vpk
CAT II 100 Vrms, 400 Vpk
with 10073C 10:1 probe: CAT I 500 Vpk, CAT II 400 Vpk
5 Vrms with 50-ohm input
±5 V on ranges <10 mV/div
±20 V on ranges 10 mV/div to 200 mV/div
±75 V on ranges >200 mV/div
Lesser of ±8 div or ±32 V
1 M
±1%, 50-ohm selectable
~ 13 pF
ac, dc
~ 25 MHz selectable
DC to max bandwidth > 40 dB
10:1 10073C shipped standard for each analog channel
Auto probe sense