Agilent Technologies 54641A Service Manual page 23

Oscilloscopes and mixed-signal oscilloscopes
Hide thumbs Also See for 54641A:
Table of Contents

Advertisement

* Denotes Warranted Specifications, all others are typical. Specifications are valid after a 30-minute warm-up period and
±10 °C from firmware calibration temperature.
General Characteristics
Physical:
Size
Weight
Calibrator Output
Trigger Out
Printer Power
Kensington lock
Power Requirements
Line Voltage Range
Line Frequency
Power Usage
Environmental Characteristics
Ambient Temperature
Humidity
Altitude
Vibration
Shock
Pollution degree2
Indoor use only
Installation categories
32.26 cm wide x 17.27 cm high x 31.75 cm deep (without handle)
6.82 kgs (15 lbs)
Frequency ~2 kHz; Amplitude ~5 V
0 to 5 V with 50
source impedance; delay ~ 22 ns
7.2 to 9.2 V, 1 A
Connection on rear panel for security
100 - 240 VAC, 50/60 Hz, CAT II, automatic selection
100 - 132 VAC, 440 Hz, CAT II, automatic selection
50/60 Hz, 100 - 240 VAC
440 Hz, 100 - 132 VAC
110 W max
Operating -10 °C to +55 °C
Non-operating -51 °C to +71 °C
Operating 95% RH at 40 °C for 24 hr
Non-operating 90% RH at 65 °C for 24 hr
Operating to 4,570 m (15,000 ft)
Non-operating to 15,244 m (50,000 ft)
HP/Agilent class B1 and MIL-PRF-28800F Class 3 random
HP/Agilent class B1 and MIL-PRF-28800F (operating 30 g, 1/2 sine, 11-ms duration, 3
shocks/axis along major axis. Total of 18 shocks)
Normally only dry non-conductive pollution occurs. Occasionally a temporary
conductivity caused by condensation must be expected.
This instrument is rated for indoor use only
CAT I: Mains isolated
CAT II: Line voltage in appliance and to wall outlet
General Information
General Characteristics

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

54642a54641d54642d

Table of Contents