Agilent Technologies 54641A Service Manual page 55

Oscilloscopes and mixed-signal oscilloscopes
Hide thumbs Also See for 54641A:
Table of Contents

Advertisement

Table 3-5
Settings Used to Verify Voltage Measurement Accuracy
Volts/Div Setting Power Supply Setting Test Limits
5 V/Div
35 V
2 V/Div
14 V
1 V/Div
7 V
0.5 V/Div
3.5 V
0.2 V/Div
1.4 V
0.1 V/Div
700 mV
50 mV/Div
350 mV
20 mV/Div
140 mV
10 mV/Div
70 mV
5 mV/Div
35 mV
2 mV/Div*
14 mV
*Full scale is defined as 32 mV on the 2 mV/div range.
Full scale on all other ranges is defined as 8 divisions times the V/div setting.
Acquire
Cursors
Y
To verify voltage measurement accuracy
34.04 V
to 35.96 V
13.616 V
to 14.384 V
6.808 V
to 7.192 V
3.404 V
to 3.596 V
1.3616 V
to 1.4384 V
680.8 mV
to 719.2 mV
340.4 mV
to 359.6 mV
136.16 mV to 143.84 mV
68.08 mV
to 71.92 mV
34.04 mV
to 35.96 mV
13.232 mV to 14.768 mV
Averaging
Mode
Normal
Y1
Testing Performance
#Avgs
X Y