Table 3-4
Testing Performance
To verify voltage measurement accuracy
Equipment Required to Verify Voltage Measurement Accuracy
Equipment
Critical Specifications
Power supply
14 mV to 35 Vdc,
0.1 mV resolution
Digital multimeter
Better than 0.01% accuracy
Cable
BNC, Qty 2
Shorting cap
BNC
Adapter
BNC (f) to banana (m)
Adapter
BNC tee (m) (f) (f)
Blocking capacitor
Recommended Model/Part
Fluke 5820A or Agilent 3245A
Agilent 34401A
Agilent 10503A
Agilent 1250-0774
Agilent 1251-2277
Agilent 1250-0781
Agilent 10240B