NurseCall N46 Wall Transmitter
en User Manual

Advertisement

Table of Contents
loading

Summary of Contents for Bosch NurseCall N46

  • Page 1 NurseCall N46 Wall Transmitter en User Manual...
  • Page 3: Table Of Contents

    NurseCall N46 Wall Transmitter | en Table of Contents Identification Document Customer support addresses Generalities NurseCall System NurseCall N46 Wall Transmitter Safety Instructions Introduction 3.1.1 Principle 3.1.2 Importance of safety instructions 3.1.3 Disregarding safety rules Environmental conditions General safety instructions 3.3.1...
  • Page 4 | NurseCall N46 Wall Transmitter Wired Inputs General configuration Call for Help Call for Assistance Call Acknowledgement (Staff Presence) Call for Help 2 (reserve) Technical Alarm Caregiver Arrival Caregiver Departure Daily Message and Battery Low Message Error Message Operation...
  • Page 5: Identification

    NurseCall N46 Wall Transmitter Identification | en Identification Document Name User Manual 970.030 Table 1.1 Document No. Version Description v1.0 | 2009.06 First Edition Table 1.2 Version Management Bosch Security Systems User Manual 970.030 | v1.0 | 2009.06...
  • Page 6: Customer Support Addresses

    | Identification NurseCall N46 Wall Transmitter Customer support addresses TeleAlarm SA Bosch Group Bosch Security Systems BV Rue du Pont 23 Postbus 80002 CH-2300 La Chaux-de-Fonds NL-5600 JB Eindhoven Switzerland Netherlands Phone: +41 32 327 25 40 Phone: +31 (0)900 8212499...
  • Page 7: Generalities

    It is part of a NurseCall system, and is easily expandable. In the NurseCall system, the NurseCall N46 Wall Transmitter allows the resident of a home or the patient of a clinic to easily alert the care personnel if needed. The patient simply has to push the large red button.
  • Page 8: Safety Instructions

    Environmental conditions WARNING! The NurseCall N46 Wall Transmitter must not be located near a water tap or any other source of water. The electrical safety of the NurseCall N46 Wall Transmitter is only guaranteed if the electrical installation is in accordance with the national regulations and if this installation works prop- erly.
  • Page 9: General Safety Instructions

    NurseCall N46 Wall Transmitter Safety Instructions | en Install the NurseCall N46 Wall Transmitter in a dry place, away from any source of heat. CAUTION! Interferences Avoid immediate proximity to other electric devices such as a television. General safety instructions...
  • Page 10: Observation And Information

    NurseCall N46 Wall Transmitter CAUTION! Never let any liquid enter the system. In case of liquid spill inside the NurseCall N46 Wall Transmitter, act immediately as follows: Switch off the NurseCall N46 Wall Transmitter by taking out the battery or removing the power supply.
  • Page 11: Description

    NurseCall N46 Wall Transmitter Description | en Description General description 4.1.1 Top view Fig. 4.1 Top view Red emergency call button LED indicator for alarm LED indicator for presence Green acknowledgment button, to acknowledge an alarm locally. Blue assistance call button...
  • Page 12: Main Functions

    | Description NurseCall N46 Wall Transmitter Main functions The main function of the NurseCall N46 Wall Transmitter is to enable easier handling of emergencies and help, directly from the patient’s room. 4.2.1 Red emergency call button Its large red button with textured surface makes it easy to identify and to press, launching an emergency call.
  • Page 13: Installation

    Store the packaging material for further use (storage or transport). NOTICE! In case of defective or missing equipment, do not try to install the NurseCall N46 Wall Trans- mitter. Contact immediately your local representative. Take all components out of the box and place the NurseCall N46 Wall Transmitter on the working space.
  • Page 14: Installation

    5.2.2 Wall installation The NurseCall N46 Wall Transmitter can be fixed with two to five screws on a wall. Two holes at 60 mm interval are provided for direct mounting on connection boxes. The three upper holes are used for positioning the device and cannot be reached from the installation compartment.
  • Page 15: Connecting To The Power Supply

    5.2.3 Connecting to the power supply The NurseCall N46 Wall Transmitter can be powered by a lithium battery or/and an external 12-24 VDC power supply. If the device is powered with an external power supply, the battery is not necessary, but it can be used as a backup power supply in case of power loss.
  • Page 16: Replacing The Battery

    Replacing the battery WARNING! – May explode if exposed to fire; – Use only original batteries intended for use with your NurseCall N46 Wall Transmitter. Using other type of batteries could be dangerous; – Do not expose the battery to liquids; –...
  • Page 17: Configuration And Wiring

    NurseCall N46 Wall Transmitter Installation | en 5.2.8 Configuration and Wiring Solder Bridges (if soldered): B8 Call for Assistance only by Staff Presence Call for Assistance from terminals J6[4-5] B6 No Staff Presence processing B5 No Daily Message B4 No calling message repetition...
  • Page 18: Inputs And Outputs

    | Installation NurseCall N46 Wall Transmitter Inputs and Outputs General inputs The * marks a factory setting. External trigger J4 Jack 3.5 mm Closing contact causes a Call for Help Contact opening causes a Call for Help Power supply...
  • Page 19: General Configuration

    The Assistance Call is possible only if the Care Personnel has signaled his/her presence. NurseCall N46 Wall Transmitters that feature the output relays are delivered with the Staff Presence Processing activated (B6 open). Models that do not feature the output relays are delivered with the Staff Presence Processing disabled (B6 soldered).
  • Page 20: Use

    | Use NurseCall N46 Wall Transmitter Every time a NurseCall N46 Wall Transmitter is used by pressing a button or activating a wired input, the device sends a radio signal and the lamps indicates its actual state. Optionally, two relays allow the transmission of status to a remote display such as the signal lamps placed on top of the door.
  • Page 21: Call Acknowledgement (Staff Presence)

    NurseCall N46 Wall Transmitter Use | en Call Acknowledgement (Staff Presence) The call acknowledgement is triggered by pushing the Acknowledgement button (green). The call acknowledgement stops immediately all repetitions of the calls. Optionally, a call acknowledgement triggers the Staff Presence signal: –...
  • Page 22: Caregiver Departure

    | Use NurseCall N46 Wall Transmitter Insertion of the coded key is signaled by green flashes on the signal lamp and by activating the Staff Presence relay. Caregiver Departure The Caregiver Departure message is triggered when the Caregiver removes the coded key from the housing, when leaving the room.
  • Page 23: Operation

    NurseCall N46 Wall Transmitter Operation | en Operation Description of the six different states During operation, the NurseCall N46 Wall Transmitter can be set to one of six following states: State Description Stand-by No call is being processed. Alarm without staff presence The patient has triggered a Call for Help but the Care Personnel has not yet responded.
  • Page 24: Legend Of Illustrations

    Call for Assistance staff presence ** ** If the function Assistance only in case of Staff Presence is activated and the NurseCall N46 Wall Transmitter is not in Staff presence state, the Call for Assistance causes a state Alarm without Staff Presence and a Call for Help radio message.
  • Page 25 Call for Assistance staff presence ** **If the function Assistance only in case of Staff Presence is activated and the NurseCall N46 Wall Transmitter is not in Staff presence state, the Call for Assistance causes a state Alarm without Staff Presence and a Call for Help radio message.
  • Page 26 | Operation NurseCall N46 Wall Transmitter State before action: Assistance with staff presence Signal lamp Radio Message Action State Relay contact Assistance with staff (none) presence Assistance with staff Call for Help presence Call for Assistance (unchanged) Call Alarm with staff...
  • Page 27: Device Behavior During Coded Key Insertion And Removal

    Staff departure with Coded key removal Stand-by *** code *** If the call has not been acknowledged, removing the key will turn the NurseCall N46 Wall Transmitter in Alarm without Staff Presence state. Bosch Security Systems User Manual 970.030 | v1.0 | 2009.06...
  • Page 28 | Operation NurseCall N46 Wall Transmitter State before action: Assistance with staff presence Signal lamp Radio Message Action State Relay contact Assistance with staff (none) presence Assistance with staff Staff arrival with Coded key insertion presence code (unchanged) Staff departure with...
  • Page 29: Operation With The Staff Presence Processing Deactivated

    NurseCall N46 Wall Transmitter Operation | en Operation with the staff presence processing deactivated 7.4.1 Device behavior upon pressing a button State before action: Stand-by Signal lamp Radio Message Action State Relay contact (none) Stand-by Alarm without staff Call for Help...
  • Page 30: Device Behavior During Coded Key Insertion And Removal

    | Operation NurseCall N46 Wall Transmitter 7.4.2 Device behavior during coded key insertion and removal NOTICE! The coded key insertion an removal functions are optional. State before action: Stand-by Action State Relay contact Signal lamp Radio Message (none) Stand-by...
  • Page 31: Storage

    Storage One can distinguish between long term storage and short term storage. If the NurseCall N46 Wall Transmitter must be temporarily stored, it may not be necessary to wrap it completely. However, if you wish to store the NurseCall N46 Wall Transmitter for a longer time, for instance in a storage room or any similar location, it is recommended to use original packing material.
  • Page 32: Disposal

    The nearest disposal locations are established in accordance with the currently applicable laws. Ask the local authorities. Returning to the manufacturer If there is no practical disposal place, the NurseCall N46 Wall Transmitter may be returned to your local representative. Materials The NurseCall N46 Wall Transmitter must be returned to an authorized point of recycling.
  • Page 33: Appendix

    NurseCall N46 Wall Transmitter Appendix | en Appendix 10.1 Electrical specifications External power supply 12-24 VDC & consumption Transmission: < 40 mA Idle mode: < 10 mA Battery power supply 3V lithium battery CR2450 & consumption Transmission: < 30 mA Idle mode: <...
  • Page 34: Ec-Declaration Of Conformity